SUB-MICROLITER FLOW-THROUGH CUVETTE FOR FLUORESCENCE MONITORING OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC EFFLUENTS

被引:106
作者
HERSHBERGER, LW [1 ]
CALLIS, JB [1 ]
CHRISTIAN, GD [1 ]
机构
[1] UNIV WASHINGTON,DEPT CHEM,SEATTLE,WA 98195
关键词
D O I
10.1021/ac50045a021
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学]; 081704 [应用化学];
摘要
A new sub-microliter flow-through cuvette suitable for monitoring liquid chromatographic effluents has been developed and tested. The cuvette is based on the sheath flow principle, in which the chromatographic effluent is injected into the center of an ensheathing solvent stream but does not mix with it because laminar flow conditions are maintained. The optical volume of the cuvette is easily varied by adjusting the relative flow rates of sheath and sample. The cuvette has been tested for stability, reproducibility, and dynamic range using porphyrins as test compounds. The detection limit for mesoporphyrin IX dimethyl ester was 8 X 10-10 M when the cuvette was operated with an effective detection volume of 53 nL. © 1979, American Chemical Society. All rights reserved.
引用
收藏
页码:1444 / 1446
页数:3
相关论文
共 11 条
[1]
A DEVICE FOR COUNTING SMALL PARTICLES SUSPENDED IN A FLUID THROUGH A TUBE [J].
CROSLANDTAYLOR, PJ .
NATURE, 1953, 171 (4340) :37-38
[2]
LASER FLUORIMETRY - SUBPICOGRAM DETECTION OF AFLATOXINS USING HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
DIEBOLD, GJ ;
ZARE, RN .
SCIENCE, 1977, 196 (4297) :1439-1441
[3]
FALK JE, 1964, PORPHYRINS METALLOPO, P233
[4]
PORPHYRIN FREE BASE PHOSPHORESCENCE [J].
GOUTERMAN, M ;
KHALIL, G .
JOURNAL OF MOLECULAR SPECTROSCOPY, 1974, 53 (01) :88-100
[5]
STUDY OF MICRO-HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .1. DEVELOPMENT OF TECHNIQUE FOR MINIATURIZATION OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ISHII, D ;
ASAI, K ;
HIBI, K ;
JONOKUCHI, T ;
NAGAYA, M .
JOURNAL OF CHROMATOGRAPHY, 1977, 144 (02) :157-168
[6]
SAMPLING AND EXTRA-COLUMN EFFECTS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - INFLUENCE OF PEAK SKEW ON PLATE COUNT CALCULATIONS [J].
KIRKLAND, JJ ;
YAU, WW ;
STOKLOSA, HJ ;
DILKS, CH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (08) :303-316
[7]
KINETIC CONDITIONS FOR OPTIMUM SPEED AND RESOLUTION IN COLUMN CHROMATOGRAPHY [J].
KNOX, JH ;
SALEEM, M .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1969, 7 (10) :614-&
[8]
PRACTICAL ASPECTS OF LC THEORY [J].
KNOX, JH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (09) :352-364
[9]
STUDY OF PERTINENCY OF PRESSURE IN LIQUID-CHROMATOGRAPHY .2. PROBLEMS IN EQUIPMENT DESIGN [J].
MARTIN, M ;
EON, C ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1975, 108 (02) :229-241
[10]
PACKED MICROCAPILLARY COLUMNS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
TSUDA, T ;
NOVOTNY, M .
ANALYTICAL CHEMISTRY, 1978, 50 (02) :271-275