COMPARISON BETWEEN INFRARED AND ULTRAVIOLET-LASER ABLATION AT ATMOSPHERIC-PRESSURE - IMPLICATIONS FOR SOLID SAMPLING INDUCTIVELY-COUPLED PLASMA SPECTROMETRY

被引:119
作者
GEERTSEN, C
BRIAND, A
CHARTIER, F
LACOUR, JL
MAUCHIEN, P
SJOSTROM, S
MERMET, JM
机构
[1] CEA,CTR ETUD SACLAY,SPECTROSCOPIE LASER LAB,SERV PHYS EXPT & ANAL,BAT 391,F-91191 GIF SUR YVETTE,FRANCE
[2] UNIV C BERNARD LYON 1,SCI ANALYT LAB,BAT 308,F-69622 VILLEURBANNE,FRANCE
关键词
LASER-PRODUCED PLASMA; SOLID SAMPLING; LASER ABLATION; INDUCTIVELY COUPLED PLASMA SPECTROMETRY; TRACE ANALYSIS;
D O I
10.1039/ja9940900017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The efficiency of laser solid sampling was investigated as a function of several experimental parameters under experimental conditions similar to those used for laser ablation (LA) inductively coupled plasma spectrometry. This was done by studying the amount of material removed as a function of melting temperature of the (metallic) matrix material, laser wavelength and laser energy, and by studying the plasma ignition in air and argon buffer gases as a function of laser wavelength. It was found that direct LA is the major process responsible for the removal of material in the case of a UV laser, as opposed to with an IR laser, where shielding of the laser radiation by the absorbing plasma limits direct LA and increases the temperature of the plasma. The consequences of this difference between IR and UV laser radiation are considerable and lead to a superior performance of UV laser sampling in every analytical aspect: reproducibility, matrix effects, quantification, spatial resolution and sensitivity.