MODIFICATION OF THE POTASSIUM-IODIDE PROCEDURE FOR IMPROVED STOICHIOMETRY

被引:13
作者
LANTING, RW
机构
[1] Research Institute for Environmental Hygiene
关键词
D O I
10.1016/0004-6981(79)90150-1
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Sample humidity and sample flow rate dependence on the stoichiometry of the 1% neutral buffered potassium iodide method for ozone calibration can be eliminated by adding potassium bromide (KBr) along with the reduction of the phosphate buffer concentration. © 1979.