MODIFIED LEAD ATTENUATION METHOD TO DETERMINE THE FAST-NEUTRON SENSITIVITY KU OF A PHOTON DOSIMETER

被引:13
作者
HOUGH, JH [1 ]
机构
[1] HAMMERSMITH HOSP,MRC,CYCLOTRON UNIT,LONDON W12 0HS,ENGLAND
关键词
D O I
10.1088/0031-9155/24/4/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Various lead attenuation techniques to determine the fast neutron sensitivity ku of a photon dosemeter are reviewed and a modified method is proposed to determine the ku value of a commercially available Geiger-Muller detector. The dependence on the 'effective' photon energy of the gamma -component of the mixed n- Gamma field is illustrated and the most probable photon energy is deduced from measurements outside the radiation field. Using the relevant photon mass attenuation coefficient, measurements carried out with a well-collimated neutron beam at the MRC Cyclotron, Hammersmith Hospital, London, show that the Dg component contributes 2.41+or-0.13% to the total radiation field. Using a detector with a PTFE sleeve in the place of a Perspex sleeve, highly consistent results are obtained and the enhancement of the ku value by the Perspex is demonstrated.
引用
收藏
页码:734 / 747
页数:14
相关论文
共 18 条
[1]  
ATTIX FH, 1974, EUR J CANCER, V10, P314
[2]  
ATTIX FH, 1977, C001105252 ERDA REP
[3]  
ATTIX FH, 1977, C001105251 ERDA REP
[4]  
ATTIX FH, 1975, 2ND P S NEUTR DOS BI, P329
[5]   FAST NEUTRON BEAM FROM MRC CYCLOTRON [J].
BEWLEY, DK ;
PARNELL, CJ .
BRITISH JOURNAL OF RADIOLOGY, 1969, 42 (496) :281-&
[6]  
CLELAND MR, 1975, 2ND P S NEUTR DOS BI, P397
[7]  
COLVETT RD, 1974, C0032433 USAEC REP, P152
[8]  
HESS A, 1978, 3RD P S NEUTR DOS BI, P181
[9]  
ITO A, 1978, 3RD P S NEUTR DOS BI, P605
[10]  
KUCHNIR FT, 1975, BIOMEDICAL DOSIMETRY, P104