INTERLABORATORY COMPARISON OF THERMOSPRAY AND PARTICLE BEAM LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY INTERFACES - EVALUATION OF A CHLORINATED PHENOXY ACID HERBICIDE LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY ANALYSIS METHOD

被引:37
作者
JONES, TL
BETOWSKI, LD
LESNIK, B
CHIANG, TC
TEBERG, JE
机构
[1] US EPA,ENVIRONM MONITORING SYST LAB,POB 93478,LAS VEGAS,NV 89193
[2] US DOE,LAB MANAGEMENT BRANCH,GERMANTOWN,MD 20874
[3] LOCKHEED ENGN & SCI CO,LAS VEGAS,NV 89119
关键词
D O I
10.1021/es00023a006
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Seven laboratories participated in an interlaboratory evaluation of a liquid chromatography/mass spectrometry (LC/MS) method for the analysis of 10 chlorinated phenoxy acid herbicides. The focus of this evaluation was to test the intercomparability of LC/MS data obtained from two types of LC/MS interfaces [i.e., thermospray (TS) and particle beam (PB)]. Eight simulated sample extracts were sent to each laboratory for LC/MS analysis. There were statistically significant differences between interfaces in the quantitative data for all analytes except 2-(2,4,5-trichlorophenoxy)propanoic acid (silvex). Particle beam exhibited a high positive bias and a low relative standard deviation at the highest sample extract concentration range, 500-mu-g/mL, while TS showed a low bias and a low relative standard deviation at the lowest sample extract concentration range, 5-mu-g/mL. Another factor of this study was to look for any performance differences between interfaces of the same type, but differing manufacturers. A statistical difference was observed, between PB interfaces, for 2-(1-methylpropyl)-4,6-dinitrophenol (dinoseb).
引用
收藏
页码:1880 / 1884
页数:5
相关论文
共 10 条