MR IMAGING OF SILICONE BREAST IMPLANTS - COMPARISON OF DIFFERENT COIL ARRAYS

被引:10
作者
SINHA, S [1 ]
GORCZYCA, DP [1 ]
DEBRUHL, ND [1 ]
SHELLOCK, FG [1 ]
GAUSCHE, VR [1 ]
BASSETT, LW [1 ]
机构
[1] UNIV CALIF LOS ANGELES,MED CTR,JONSSON COMPREHENS CANC CTR,LOS ANGELES,CA 90024
关键词
BREAST; PROSTHESES; MAGNETIC RESONANCE (MR); COIL ARRAYS; SILICONE;
D O I
10.1148/radiology.187.1.8451430
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To determine the optimal coil and patient position for magnetic resonance imaging of patients with silicone breast implants, images obtained with body, circular surface, flexible, single shoulder, and dual shoulder coils in various numbers of patients were compared with regard to signal-to-noise ratio (S/N), decay of signal intensity, resolution, and key area coverage. Images obtained with the dual shoulder coil and prone position, with a customized positioning device, gave the best S/N, resolution, and coverage of the breasts and axillae.
引用
收藏
页码:284 / 286
页数:3
相关论文
共 4 条
  • [1] IMPROVED IMAGING OF THE AUGMENTED BREAST
    EKLUND, GW
    BUSBY, RC
    MILLER, SH
    JOB, JS
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (03) : 469 - 473
  • [2] AUGMENTATION MAMMAPALSTY - NORMAL AND ABNORMAL FINDINGS WITH MAMMOGRAPHY AND ULTRASOUND
    GANOTT, MA
    HARRIS, KM
    ILKHANIPOUR, ZS
    COSTAGRECO, MA
    [J]. RADIOGRAPHICS, 1992, 12 (02) : 281 - 295
  • [3] SILICONE BREAST IMPLANTS INVIVO - MR IMAGING
    GORCZYCA, DP
    SINHA, S
    AHN, CY
    DEBRUHL, ND
    HAYES, MK
    GAUSCHE, VR
    SHAW, WW
    BASSETT, LW
    [J]. RADIOLOGY, 1992, 185 (02) : 407 - 410
  • [4] DEFINITIVE DIAGNOSIS OF BREAST IMPLANT RUPTURE BY ULTRASONOGRAPHY
    LEVINE, RA
    COLLINS, TL
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 1991, 87 (06) : 1126 - 1128