DEVELOPMENT OF 180-GHZ HETERODYNE RADIOMETER FOR ELECTRON-CYCLOTRON EMISSION MEASUREMENTS IN JT-60U

被引:26
作者
ISEI, N
SATO, M
ISHIDA, S
UCHINO, K
NAGASHIMA, A
MATOBA, T
OYEVAAR, T
机构
[1] Japan Atomic Energy Research Institute, Naka Fusion Research Establishment, Naka-gun, Ibaraki-ken 311-01, Naka-machi
关键词
D O I
10.1063/1.1146364
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 12-channel heterodyne radiometer system has been developed to obtain electron temperature with high time and spatial resolution from the electron cyclotron emission (ECE) in JT-60U. Resolutions of a single-sideband-type receiver are (1) Δt∼1 μs and (2) Δf=0.5 GHz. The second harmonic extraordinary mode of ECE is utilized for this system and the frequency band is 176-188 GHz. The noise level of the system is determined by unavoidable photon fluctuations, and the signal-to-noise ratio was measured to be about 50 with a video bandwidth of 100 kHz for a signal temperature above ∼200 eV. Details of the design, the obtained performance, and typical measurements are described. © 1995 American Institute of Physics.
引用
收藏
页码:413 / 415
页数:3
相关论文
共 10 条
  • [1] MEASUREMENT OF APPARENT TURBULENT TEMPERATURE-FLUCTUATIONS ON THE TFR TOKAMAK
    CAVALLO, A
    CANO, R
    [J]. PLASMA PHYSICS AND CONTROLLED FUSION, 1981, 23 (01) : 61 - 65
  • [2] COSTLEY A. E., 1982, P COURSE DIAGNOSTIC, VI, P129
  • [3] HARTFUS HJ, 1989, EC7 WORKSHOP ECE ECR, P253
  • [4] HARTFUS HJ, 1989, EC7 WORKSHOP ECE ECR, P267
  • [5] ISHIDA S, 1990, REV SCI INSTRUM, V61, P2836
  • [6] SALMON NA, 1988, THESIS U LONDON
  • [7] SATO M, 1993, JAERIM93057, P359
  • [8] VERRECK M, 1989, EC7 WORKSHOP ECE ECR, P322
  • [9] WAIDMANN G, NUCL FUSION, V32, P645
  • [10] WAIDMANN G, 1989, EC7 WORKSHOP ECE ECR, P92