IMMUNOCHEMICAL ASSAYS OF SERUM-PROTEINS - A EUROPEAN EXTERNAL QUALITY ASSESSMENT SURVEY AND THE EFFECTS OF CALIBRATION PROCEDURES ON INTERLABORATORY AGREEMENT

被引:34
作者
BULLOCK, DG
DUMONT, G
VASSAULT, A
AGUZZI, F
CHAMBERS, RE
WARD, AM
WHICHER, JT
BIENVENU, J
机构
[1] UNIV LEEDS,SCH MED,DEPT CHEM PATHOL,LEEDS LS2 9JT,W YORKSHIRE,ENGLAND
[2] QUEEN ELIZABETH MED CTR,DEPT CLIN CHEM,UKEQAS GEN CLIN CHEM,BIRMINGHAM,ENGLAND
[3] HOP NECKER ENFANTS MALAD,BIOCHIM LAB A,F-75730 PARIS 15,FRANCE
[4] OSPED STRADELLA,ANAL LAB,STRADELLA,ITALY
[5] BRISTOL ROYAL INFIRM & GEN HOSP,DEPT CHEM PATHOL,BRISTOL BS2 8HW,AVON,ENGLAND
[6] ROYAL HALLAMSHIRE HOSP,DEPT IMMUNOL,SHEFFIELD S10 2JF,S YORKSHIRE,ENGLAND
[7] CTR HOSP LYON SUD,BIOCHIM LAB,F-69310 PIERRE BENITE,FRANCE
关键词
Calibration; External quality assessment; Immunochemical protein assay; Interlaboratory agreement; Reference material;
D O I
10.1016/0009-8981(90)90258-T
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
An external quality assessment survey of immunochemical assays of 9 proteins (immunoglobulins G, A and M, complement components C3 and C4, alpha1 -antitrypsin, orosomucoid, haptoglobin and transferrin) in 5 European countries (Austria, France, Hungary, Italy and UK) showed inter-country differences in the mean values obtained. Reprocessing of the results using one of the two specimens distributed as a 'calibrant' effectively eliminated or reduced substantially these differences. Consideration of the methods used by participants confirmed previous indications from national surveys that the differences were due to lack of agreement among commercial calibrants. Such interlaboratory variations were also minimised by the 'calibration' in this survey. The role of European working calibration materials in ensuring interlaboratory agreement on an international basis is discussed. © 1990.
引用
收藏
页码:21 / 35
页数:15
相关论文
共 17 条
[1]  
[Anonymous], 1981, EXTERNAL QUALITY ASS
[2]   EXTERNAL QUALITY ASSESSMENT OF URINARY PREGNANCY ESTROGEN ASSAY - FURTHER EXPERIENCE IN THE UNITED-KINGDOM [J].
BULLOCK, DG ;
WILDE, CE .
ANNALS OF CLINICAL BIOCHEMISTRY, 1985, 22 (MAY) :273-282
[3]   EXTERNAL QUALITY ASSESSMENT OF SERUM ENZYME-ACTIVITY ASSAYS AND THE EFFECT OF CALIBRATION ON INTERLABORATORY CONCORDANCE [J].
BULLOCK, DG ;
MOSS, DW ;
WHITEHEAD, TP .
ANNALS OF CLINICAL BIOCHEMISTRY, 1986, 23 :577-584
[4]   EXTERNAL QUALITY ASSESSMENT OF IMMUNOASSAYS FOR SPECIFIC PROTEINS IN SERUM - 18 MONTHS EXPERIENCE IN THE UNITED-KINGDOM [J].
CHAMBERS, RE ;
WHICHER, JT ;
BULLOCK, DG .
ANNALS OF CLINICAL BIOCHEMISTRY, 1984, 21 (JUL) :246-253
[5]   IMPROVED BETWEEN-LABORATORY AGREEMENT FOR SPECIFIC PROTEIN ASSAYS IN SERUM FOLLOWING INTRODUCTION OF A COMMON REFERENCE PREPARATION (SPS-01) DEMONSTRATED IN AN EXTERNAL QUALITY ASSESSMENT SCHEME [J].
CHAMBERS, RE ;
BULLOCK, DG ;
WHICHER, JT .
CLINICA CHIMICA ACTA, 1987, 164 (02) :189-200
[6]  
REIMER CB, 1976, CLIN CHEM, V22, P577
[7]   COLLABORATIVE CALIBRATION OF 3 ESTABLISHED REFERENCE PREPARATIONS FOR SPECIFIC PROTEINS IN HUMAN-SERA AS SECONDARY STANDARDS FOR IMMUNOGLOBULIN-A, IMMUNOGLOBULIN-M, AND IMMUNOGLOBULIN-G [J].
REIMER, CB ;
SMITH, SJ ;
WELLS, TW .
JOURNAL OF BIOLOGICAL STANDARDIZATION, 1981, 9 (04) :393-400
[8]  
REIMER CB, 1982, AM J CLIN PATHOL, V77, P12
[9]  
RITCHIE RF, 1982, AM J CLIN PATHOL, V78, P644
[10]  
ROWE DS, 1970, B WORLD HEALTH ORGAN, V42, P535