COMPARISON OF 2 DIFFERENT CELL-CULTURE METHODS IN EVALUATION OF BIOCOMPATIBILITY OF LATEX URINARY CATHETERS

被引:6
作者
KORHONEN, P
TALJA, M
RUUTU, M
ANDERSSON, LC
ALFTHAN, O
机构
[1] PAIJAT HAME CENT HOSP,SF-15850 LAHTI,FINLAND
[2] KYMENLAAKSO CENT HOSP,KOTKA,FINLAND
[3] UNIV HELSINKI,CENT HOSP,DEPT SURG 4,SF-00100 HELSINKI 10,FINLAND
[4] UNIV HELSINKI,CENT HOSP,DEPT SURG 2,UROL UNIT,SF-00100 HELSINKI 10,FINLAND
[5] UNIV HELSINKI,INST PATHOL,SF-00100 HELSINKI 10,FINLAND
来源
UROLOGICAL RESEARCH | 1991年 / 19卷 / 02期
关键词
URINARY CATHETERS; BIOCOMPATIBILITY TESTING; CELL CULTURE; TOXICITY;
D O I
10.1007/BF00368190
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
International biocompatibility standards for urinary catheters have not yet been set. The used in vivo animal tests earlier have been shown to be insensitive and expensive. The present study compared the sensitivity and other properties of two in vitro cell culture methods, the reference method of the British Standards Institution (BSI) [2] and the thymidine incorporation method (DNA synthesis inhibition test, [7]. The cell culture toxicities of thirty-seven latex catheters were measured and 84.8% of the catheters tested passed the BSI reference test while 94.6% passed the thymidine incorporation test. The overall batch-to-batch correlation was poor (p > 0.05), but within brands the correlation was better. This is obviously to be explained by the differences the characteristics of the tests and/or the chemical composition of the catheters. The thymidine incorporation test is easier to perform, requires less manpower, and is therefore less liable to subjective interpretations than the BSI test. The biocompatibility toxicity limits for latex urinary catheters need to be tightened. Our experience in this study indicates that the thymidine incorporation test or tests similar to it, which can be highly automated can be recommended for biocompatibility screening in large series, and BSI reference test can be used additionally in unclear cases.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 12 条
[1]   CATHETERS AND URETHRAL STRICTURES [J].
BLACKLOCK, NJ .
BRITISH JOURNAL OF UROLOGY, 1986, 58 (05) :475-478
[2]   VOLATILE N-NITROSAMINES IN URINARY CATHETERS [J].
HEENAN, MP ;
NACEY, JN ;
DELAHUNT, B ;
FERGUSON, AF ;
DICKSON, SJ .
BRITISH JOURNAL OF UROLOGY, 1989, 63 (01) :72-75
[3]   THE ASSESSMENT OF CATHETER-INDUCED URETHRITIS USING AN EXPERIMENTAL DOG-MODEL [J].
NACEY, JN ;
DELAHUNT, B ;
TULLOCH, AGS .
JOURNAL OF UROLOGY, 1985, 134 (03) :623-625
[4]   THE ASSESSMENT OF URINARY CATHETER TOXICITY USING CELL-CULTURES - VALIDATION BY COMPARISON WITH AN ANIMAL-MODEL [J].
NACEY, JN ;
HORSFALL, DJ ;
DELAHUNT, B ;
MARSHALL, VR .
JOURNAL OF UROLOGY, 1986, 136 (03) :706-709
[5]   CYTO-TOXICITY OF LATEX URINARY CATHETERS [J].
RUUTU, M ;
ALFTHAN, O ;
TALJA, M ;
ANDERSSON, LC .
BRITISH JOURNAL OF UROLOGY, 1985, 57 (01) :82-87
[6]  
RUUTU M, 1982, LANCET, V1, P218
[7]  
SMITH JM, 1982, LANCET, V1, P392
[8]  
SUTHERLAND PD, 1982, LANCET, V1, P623
[9]   TOXICITY TESTING OF URINARY CATHETERS [J].
TALJA, M ;
ANDERSSON, LC ;
RUUTU, M ;
ALFTHAN, O .
BRITISH JOURNAL OF UROLOGY, 1985, 57 (05) :579-584
[10]  
WALSH A, 1982, LANCET, V1, P392