CLOSED-CYCLE HELIUM CRYOSTAT MOUNT FOR THE HUBER 511.1 DIFFRACTOMETER CIRCLE

被引:14
作者
GRAAFSMA, H
SAGERMAN, G
COPPENS, P
机构
[1] State Univ of New York at Buffalo, Buffalo
关键词
D O I
10.1107/S0021889891003485
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Modifications to the high-stability mounting device for the closed-cycle helium cryostat as developed by Henriksen, Larsen & Rasmussen [J. Appl. Cryst. (1986), 19, 390-394] are described. They allow use of the cryostat on the smaller Huber 511.1 diffractometer circle with a smaller source-to-sample distance and reduced blind and collision zones.
引用
收藏
页码:961 / 962
页数:2
相关论文
共 2 条