SEPARATION OF METAL-COMPLEXES OF ETHYLENEDIAMINETETRAACETIC ACID IN ENVIRONMENTAL WATER SAMPLES BY ION CHROMATOGRAPHY WITH UV AND POTENTIOMETRIC DETECTION

被引:39
作者
BUCHBERGER, W
HADDAD, PR
ALEXANDER, PW
机构
[1] UNIV NEW S WALES,DEPT ANALYT CHEM,KENSINGTON,NSW 2033,AUSTRALIA
[2] JOHANNES KEPLER UNIV,DEPT ANALYT CHEM,A-4040 LINZ,AUSTRIA
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 558卷 / 01期
关键词
D O I
10.1016/0021-9673(91)80123-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An ion chromatographic method is described for the analysis of several metal complexes of ethylenediaminetetraacetic acid (EDTA) in water samples. Separations are achieved on C18 bonded silica, typically using a mobile phase comprising 1% (w/v) cetrimide-1.2 mM phosphate buffer (pH 7) 250% (v/v) acetonitrile-15% (v/v) methanol, as well as on a polymer-based anion exchanger using 2 mM phosphate buffer (pH 7) as eluent. Direct UV detection at 250 nm is employed for EDTA complexes of Fe(III), Cu(II), Pb(II) and Ni(II), whilst UV detection at 250 nm after post-column reaction with copper ions is utilized for EDTA complexes of Zn(II), Cd(II), Co(II), Pb(II), Ni(II) and Cu(II). Detection limits are in the range 1.5-4.0 ng for direct UV detection and 30-50 ng for post-column reaction detection. Indirect potentiometric detection after post-column reaction with copper ions is utilized with metallic copper as the indicator electrode, giving detection limits in the range 1.0-1.5-mu-g. These separations are applied to the determination of metal-EDTA complexes in river water at the ppb(b) level and to remobilization studies of metal ions in sediment.