GROSS AND MICROSCOPIC CHANGES ASSOCIATED WITH A NONTHORACOTOMY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

被引:14
作者
EPSTEIN, AE
ANDERSON, PG
KAY, GN
DAILEY, SM
PLUMB, VJ
SHEPARD, RB
机构
[1] UNIV ALABAMA,DEPT PATHOL,BIRMINGHAM,AL 35294
[2] UNIV ALABAMA,DEPT SURG,DIV CARDIOTHORAC SURG,BIRMINGHAM,AL 35294
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 04期
关键词
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; SUDDEN DEATH; VENTRICULAR TACHYCARDIA; VENTRICULAR FIBRILLATION;
D O I
10.1111/j.1540-8159.1992.tb05132.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The pathology associated with an investigational transvenous defibrillating and sensing lead is described. The lead system had delivered a total of 865 J from the time of implantation to the time of patient death from a noncardiac cause 7 months after implantation and 1 month after his last defibrillator shock. There was mild, superficial fibrous thickening on the endothelial surface of the superior vena cava adjacent to the proximal spring electrode, which did not extend into the vessel wall. The distal portion of endocardial lead was embedded in the interventricular septum near the apex of the right ventricle, surrounded by fibrous thickening, and partially covered by endocardial tissue. Microscopically, there was a thick bed of fibrous connective tissue surrounding the lead with extensive interstitial fibrous connective tissue radiating into the adjacent myocardium. Since this pattern is different from the more generalized fibrotic scarring produced by myocardial infarction, we speculate that the mechanism for the observed interstitial fibrosis is replacement fibrosis following acute myocyte injury that resulted from prior defibrillator shocks and possibly from the trauma produced by the lead compressing adjacent myocardium during systole. Potential effects on device efficacy of these fibrotic changes at the bioelectric interface include their representing a new arrhythmia substrate, the possibility that fibrosis could increase both defibrillation and pacing thresholds, and that the inflammatory reaction may cause deterioration of intracardiac electrograms and interfere with sensing and tachycardia recognition.
引用
收藏
页码:382 / 386
页数:5
相关论文
共 12 条
[1]   AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DISCHARGES AND ACUTE MYOCARDIAL INJURY [J].
AVITALL, B ;
PORT, S ;
GAL, R ;
MCKINNIE, J ;
TCHOU, P ;
JAZAYERI, M ;
TROUP, P ;
AKHTAR, M .
CIRCULATION, 1990, 81 (05) :1482-1487
[2]   REENTRY AS A CAUSE OF VENTRICULAR-TACHYCARDIA IN PATIENTS WITH CHRONIC ISCHEMIC HEART-DISEASE - ELECTROPHYSIOLOGIC AND ANATOMIC CORRELATION [J].
DEBAKKER, JMT ;
VANCAPELLE, FJL ;
JANSE, MJ ;
WILDE, AAM ;
CORONEL, R ;
BECKER, AE ;
DINGEMANS, KP ;
VANHEMEL, NM ;
HAUER, RNW .
CIRCULATION, 1988, 77 (03) :589-606
[3]   INTERNAL CARDIAC DEFIBRILLATION - HISTOPATHOLOGY AND TEMPORAL STABILITY OF DEFIBRILLATION ENERGY-REQUIREMENTS [J].
FAIN, ES ;
BILLINGHAM, M ;
WINKLE, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (03) :631-638
[4]   RECURRENT SUSTAINED VENTRICULAR-TACHYCARDIA - STRUCTURE AND ULTRASTRUCTURE OF SUBENDOCARDIAL REGIONS IN WHICH TACHYCARDIA ORIGINATES [J].
FENOGLIO, JJ ;
PHAM, TD ;
HARKEN, AH ;
HOROWITZ, LN ;
JOSEPHSON, ME ;
WIT, AL .
CIRCULATION, 1983, 68 (03) :518-533
[5]   MECHANISMS RESPONSIBLE FOR COUNTERSHOCK-INDUCED VENTRICULAR-TACHYCARDIA IN THE INTACT CANINE HEART [J].
LEJA, F ;
SCANLON, PJ ;
EULER, DE .
AMERICAN HEART JOURNAL, 1987, 113 (02) :296-301
[6]   AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR IMPLANTATION WITHOUT THORACOTOMY USING AN ENDOCARDIAL AND SUBMUSCULAR PATCH SYSTEM [J].
MCCOWAN, R ;
MALONEY, J ;
WILKOFF, B ;
SIMMONS, T ;
KHOURY, D ;
MCALISTER, H ;
MORANT, V ;
CASTLE, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (02) :415-421
[7]   MORTALITY IN PATIENTS WITH IMPLANTED AUTOMATIC DEFIBRILLATORS [J].
MIROWSKI, M ;
REID, PR ;
WINKLE, RA ;
MOWER, MM ;
WATKINS, L ;
STINSON, EB ;
GRIFFITH, LSC ;
KALLMAN, CH ;
WEISFELDT, ML .
ANNALS OF INTERNAL MEDICINE, 1983, 98 (05) :585-588
[8]   PREVENTION OF SUDDEN ARRHYTHMIC DEATH WITH IMPLANTED AUTOMATIC DEFIBRILLATORS [J].
MIROWSKI, M .
ANNALS OF INTERNAL MEDICINE, 1982, 97 (04) :606-608
[9]   PATHO-PHYSIOLOGY OF VENTRICULAR TACHYARRHYTHMIAS AMENABLE TO ELECTRIC CONTROL [J].
REID, PR ;
MOWER, MM ;
MIROWSKI, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :505-513
[10]   INITIAL CLINICAL-EXPERIENCE WITH ENDOCARDIAL DEFIBRILLATION USING AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH A TRIPLE-ELECTRODE SYSTEM [J].
SAKSENA, S ;
TULLO, NG ;
KROL, RB ;
MAURO, AM .
ARCHIVES OF INTERNAL MEDICINE, 1989, 149 (10) :2333-2339