UPPER URINARY-TRACT DETERIORATION AFTER IMPLANTATION OF ARTIFICIAL URINARY SPHINCTER

被引:14
作者
BITSCH, M
NERSTROM, H
NORDLING, J
HALD, T
机构
[1] Department of Urology, Herlev Hospital, University of Copenhagen
来源
SCANDINAVIAN JOURNAL OF UROLOGY AND NEPHROLOGY | 1990年 / 24卷 / 01期
关键词
Artificial sphincter; Myelomeningocele; Unstable bladder; Urinary incontinence; Urography;
D O I
10.3109/00365599009180356
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The preoperative urodynamic evaluations of twenty patients with myelomeningocele who had had artificial sphincter implantation because of urinary incontinence were reviewed, Four patients developed hydronephrosis and severe impairment of renal function between two and six years after implantation of the artificial sphincter, The condition was partly reversible after removal of the artificial sphincter, The urodynamic evaluation prior to implantation revealed in the four mentioned patients compared to the 16 patients with normal upper urinary tract, a tendency to lower bladder compliance, lower bladder capacity and more severe detrusor hyperre-flexia, but it was not possible to make a clear discrimination between the two groups, Attention is drawn to this unfortunate combination of effects after artificial sphincter implantation, Periodic control of the upper urinary tract by urography is recommended. © 1990 Informa UK Ltd.
引用
收藏
页码:31 / 34
页数:4
相关论文
共 9 条
  • [1] BARRETT DM, 1987, J UROLOGY, V128, P484
  • [2] DETRUSOR INSTABILITY - A DELAYED COMPLICATION IN ASSOCIATION WITH THE ARTIFICIAL SPHINCTER
    BAUER, SB
    REDA, EF
    COLODNY, AH
    RETIK, AB
    [J]. JOURNAL OF UROLOGY, 1986, 135 (06) : 1212 - 1215
  • [3] URINARY UNDIVERSION IN MYELODYSPLASIA - CRITERIA FOR SELECTION AND PREDICTIVE VALUE OF URODYNAMIC EVALUATION
    BAUER, SB
    COLODNY, AH
    HALLET, M
    KHOSHBIN, S
    RETIK, AB
    [J]. JOURNAL OF UROLOGY, 1980, 124 (01) : 89 - 93
  • [4] CARDENAS DD, 1987, ARCH PHYS MED REHAB, V68, P291
  • [5] MENINGOMYELOCELE - A CLINICAL, URODYNAMIC AND NEUROPHYSIOLOGICAL EVALUATION
    LIGHT, JK
    FAGANEL, J
    ROTH, DR
    DIMITRIJEVIC, MR
    [J]. JOURNAL OF UROLOGY, 1984, 131 (04) : 717 - 721
  • [6] ALTERATION IN DETRUSOR BEHAVIOR AND THE EFFECT ON RENAL-FUNCTION FOLLOWING INSERTION OF THE ARTIFICIAL URINARY SPHINCTER
    LIGHT, JK
    PIETRO, T
    [J]. JOURNAL OF UROLOGY, 1986, 136 (03) : 632 - 635
  • [7] URINARY-TRACT DETERIORATION ASSOCIATED WITH THE ARTIFICIAL URINARY SPHINCTER
    ROTH, DR
    VYAS, PR
    KROOVAND, RL
    PERLMUTTER, AD
    [J]. JOURNAL OF UROLOGY, 1986, 135 (03) : 528 - 530
  • [8] THE IMPACT OF THE ARTIFICIAL URINARY SPHINCTER IN THE NEUROGENIC BLADDER ON THE UPPER URINARY TRACTS
    SCOTT, FB
    FISHMAN, IJ
    SHABSIGH, R
    [J]. JOURNAL OF UROLOGY, 1986, 136 (03) : 636 - 642
  • [9] STOTT MA, 1987, SEP P INT CONT SOC B