METHODOLOGICAL PROBLEMS RELATED TO THE QUANTITATIVE ASSESSMENT OF STRETCH, ELASTIC RECOIL, AND BALLOON-ARTERY RATIO

被引:44
作者
HERMANS, WRM [1 ]
RENSING, BJ [1 ]
STRAUSS, BH [1 ]
SERRUYS, PW [1 ]
机构
[1] ERASMUS UNIV,THORAXCTR,CATHETERIZAT LAB,POB 1738,3000 DR ROTTERDAM,NETHERLANDS
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1992年 / 25卷 / 03期
关键词
PTCA; BALLOON; QCA; RECOIL;
D O I
10.1002/ccd.1810250303
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The (inflated) balloon is important to determine the extent of stretch (theoretical maximal gain in diameter or area during PTCA), elastic recoil (the loss in diameter or area immediately after PTCA), and whether under- or over-sizing (balloon-artery ratio) of the dilated lesion occurred. In these assessments, the inflated balloon is used as scaling device with assumed uniformity along its entire length. In order to assess more accurately stretch, elastic recoil, and the balloon-artery ratio, the balloon diameter was measured over its entire length with edge detection and videodensitometry in 505 lesions (453 patients). With an average inflation pressure of 8.3 +/- 2.6 atm a difference between the minimal and the maximal balloon diameter of 0.59 +/- 0.23 mm was measured using edge detection and 1.70 +/- 0.90 mm2 difference in area using videodensitometry. This results in large variations in the calculated stretch, elastic recoil, and balloon-artery ratio depending on the site of the balloon chosen for assessment. The mean difference +/- SD between stretch and elastic recoil assessed by edge detection and videodensitometry (using the minimal luminal diameter or area of the balloon) are respectively 0.00 +/- 0.19 and 0.00 +/- 0.24, suggesting that both methods are appropriate.
引用
收藏
页码:174 / 185
页数:12
相关论文
共 31 条
[1]   QUANTITATIVE AND QUALITATIVE EFFECTS OF INTRACORONARY STREPTOKINASE IN UNSTABLE ANGINA AND NON-Q-WAVE INFARCTION [J].
AMBROSE, JA ;
HJEMDAHLMONSEN, C ;
BORRICO, S ;
SHERMAN, W ;
COHEN, M ;
GORLIN, R ;
FUSTER, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (05) :1156-1165
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   IN-HOSPITAL MORBIDITY AND MORTALITY IN PATIENTS UNDERGOING ELECTIVE CORONARY ANGIOPLASTY [J].
BREDLAU, CE ;
ROUBIN, GS ;
LEIMGRUBER, PP ;
DOUGLAS, JS ;
KING, SB ;
GRUENTZIG, AR .
CIRCULATION, 1985, 72 (05) :1044-1052
[4]   QUANTITATIVE CORONARY ARTERIOGRAPHY - ESTIMATION OF DIMENSIONS, HEMODYNAMIC RESISTANCE, AND ATHEROMA MASS OF CORONARY-ARTERY LESIONS USING ARTERIOGRAM AND DIGITAL COMPUTATION [J].
BROWN, BG ;
BOLSON, E ;
FRIMER, M ;
DODGE, HT .
CIRCULATION, 1977, 55 (02) :329-337
[5]   INVIVO ASSESSMENT OF VASCULAR DILATATION DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
JAIN, A ;
DEMER, LL ;
RAIZNER, AE ;
HARTLEY, CJ ;
LEWIS, JM ;
ROBERTS, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (13) :988-992
[6]   CHANGES IN CROSS-SECTIONAL AREA OF THE CORONARY LUMEN IN THE 6 MONTHS AFTER ANGIOPLASTY - A QUANTITATIVE-ANALYSIS OF THE VARIABLE RESPONSE TO PERCUTANEOUS TRANS-LUMINAL ANGIOPLASTY [J].
JOHNSON, MR ;
BRAYDEN, GP ;
ERICKSEN, EE ;
COLLINS, SM ;
SKORTON, DJ ;
HARRISON, DG ;
MARCUS, ML ;
WHITE, CW .
CIRCULATION, 1986, 73 (03) :467-475
[7]   ASSESSMENT OF CORONARY ANGIOPLASTY BY AN AUTOMATED DIGITAL ANGIOGRAPHIC METHOD [J].
KATRITSIS, D ;
LYTHALL, DA ;
ANDERSON, MH ;
COOPER, IC ;
WEBBPEPLOE, MM .
AMERICAN HEART JOURNAL, 1988, 116 (05) :1181-1187
[8]   ASSESSMENT OF CORONARY-ARTERY STENOSES BY DIGITAL SUBTRACTION ANGIOGRAPHY - A PATHOANATOMIC VALIDATION [J].
KLEIN, LW ;
AGARWAL, JB ;
ROSENBERG, MC ;
STETS, G ;
WEINTRAUB, WS ;
SCHNEIDER, RM ;
HERMANN, G ;
HELFANT, RH .
AMERICAN HEART JOURNAL, 1987, 113 (04) :1011-1017
[9]   INFLUENCE OF INTIMAL DISSECTION ON RESTENOSIS AFTER SUCCESSFUL CORONARY ANGIOPLASTY [J].
LEIMGRUBER, PP ;
ROUBIN, GS ;
ANDERSON, HV ;
BREDLAU, CE ;
WHITWORTH, HB ;
DOUGLAS, JS ;
KING, SB ;
GREUNTZIG, AR .
CIRCULATION, 1985, 72 (03) :530-535
[10]   AUTOMATED QUANTITATIVE CORONARY ARTERIOGRAPHY - MORPHOLOGICAL AND PHYSIOLOGICAL VALIDATION IN VIVO OF A RAPID DIGITAL ANGIOGRAPHIC METHOD [J].
MANCINI, GBJ ;
SIMON, SB ;
MCGILLEM, MJ ;
LEFREE, MT ;
FRIEDMAN, HZ ;
VOGEL, RA .
CIRCULATION, 1987, 75 (02) :452-460