IMPROVED LOW-FREQUENCY STABILITY OF BOLOMETRIC DETECTORS

被引:23
作者
WILBANKS, T
DEVLIN, M
LANGE, AE
SATO, S
BEEMAN, JW
HALLER, EE
机构
[1] NAGOYA UNIV,DEPT PHYS,NAGOYA 46401,JAPAN
[2] UNIV CALIF BERKELEY LAWRENCE BERKELEY LAB,CTR ADV MAT,BERKELEY,CA 94720
[3] UNIV CALIF BERKELEY,DEPT MAT SCI,BERKELEY,CA 94720
基金
美国国家航空航天局;
关键词
D O I
10.1109/23.106678
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed an AC bridge readout system that greatly improves the low-frequency stability of bolometric detectors. The readout can be implemented with a simple circuit appropriate for use in space applications. Using a matched pair of detectors in the readout, we have achieved system noise within a factor of two of the fundamental noise limit of the detectors at frequencies as low as 10 mHz. The low frequency stability of the readout system allows slower, more sensitive detectors to be used in many applications, and facilitates observing strategies that are well-suited to space-borne observations. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:566 / 572
页数:7
相关论文
共 13 条
[1]   SUPERCONDUCTIVE BOLOMETERS FOR SUBMILLIMETER WAVELENGTHS [J].
CLARKE, J ;
HOFFER, GI ;
RICHARDS, PL ;
YEH, NH .
JOURNAL OF APPLIED PHYSICS, 1977, 48 (12) :4865-4879
[2]  
DUBAND L, 1990, IN PRESS ADV CRYOGEN
[3]  
DUBAND L, 1990, IN PRESS CRYOGENICS
[4]   PHYSICS AND DESIGN OF ADVANCED IR BOLOMETERS AND PHOTOCONDUCTORS [J].
HALLER, EE .
INFRARED PHYSICS, 1985, 25 (1-2) :257-266
[5]   IMPROVED FABRICATION TECHNIQUES FOR INFRARED BOLOMETERS [J].
LANGE, AE ;
KREYSA, E ;
MCBRIDE, SE ;
RICHARDS, PL ;
HALLER, EE .
INTERNATIONAL JOURNAL OF INFRARED AND MILLIMETER WAVES, 1983, 4 (05) :689-706
[6]   LOW-TEMPERATURE GERMANIUM BOLOMETER [J].
LOW, FJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1961, 51 (11) :1300-&
[7]   BOLOMETERS - ULTIMATE SENSITIVITY, OPTIMIZATION, AND AMPLIFIER COUPLING [J].
MATHER, JC .
APPLIED OPTICS, 1984, 23 (04) :584-588
[8]   BOLOMETER NOISE - NON-EQUILIBRIUM THEORY [J].
MATHER, JC .
APPLIED OPTICS, 1982, 21 (06) :1125-1129
[9]  
MCCAMMON D, 1987, 18TH P INT C LOW T 3, P2084
[10]   NEUTRON-TRANSMUTATION-DOPED GERMANIUM BOLOMETERS [J].
PALAIO, NP ;
RODDER, M ;
HALLER, EE ;
KREYSA, E .
INTERNATIONAL JOURNAL OF INFRARED AND MILLIMETER WAVES, 1983, 4 (06) :933-943