CORONARY BLOOD-FLOW MEASUREMENT USING AN ANGIOGRAPHIC 1ST PASS DISTRIBUTION TECHNIQUE - A FEASIBILITY STUDY

被引:20
作者
HANGIANDREOU, NJ
FOLTS, JD
PEPPLER, WW
MISTRETTA, CA
机构
[1] UNIV WISCONSIN,DEPT MED,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT RADIOL,MADISON,WI 53706
关键词
DIGITAL SUBTRACTION ANGIOGRAPHY; VIDEODENSITOMETRY; CORONARY FLOW RESERVE;
D O I
10.1118/1.596716
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to the well-documented problems associated with visual interpretation of coronary angiograms, more physiologic means of assessing coronary artery stenosis are being investigated. One physiologic parameter that has been suggested is coronary flow reserve (CFR). A digital subtraction angiographic technique based on first pass distribution analysis (FPA) is proposed as a means of measuring CFR and absolute coronary flow. The theory of the FPA method is first outlined, and the implementation of a preliminary version of the FPA algorithm is described. Experiments verifying the utility of this algorithm for measuring absolute flow through a flow phantom, and through the canine circumflex artery are reported. It was determined that the preliminary FPA algorithm is capable of measuring canine coronary flow ratios (R) with accuracy and precision characteristics meeting or exceeding those reported for the parametric imaging technique (R(FPA) = 0.933.R(true), SEE = 0.16, r = 0.984). Accurate absolute flow (Q) measurements were obtained in all of the phantom experiments (Q(FPA) = 1.054.Q(true), r = 0.993), and in one of the three dogs that were studied (Q(FPA) = Q(true), r = 0.935). The difficulty encountered in the other two dog experiments is attributed to the effects of system temporal lag, and would likely be corrected through the use of improved cameras. The feasibility of the general FPA method for measuring relative flow is established, and the potential for routine, absolute flow measurement is demonstrated.
引用
收藏
页码:947 / 954
页数:8
相关论文
共 25 条
[1]   THE CHRONIC CORONARY FLOW RESERVE PROVIDED BY SAPHENOUS-VEIN BYPASS GRAFTS AS DETERMINED BY DIGITAL CORONARY RADIOGRAPHY [J].
BATES, ER ;
VOGEL, RA ;
LEFREE, MT ;
KIRLIN, PC ;
ONEILL, WW ;
PITT, B .
AMERICAN HEART JOURNAL, 1984, 108 (03) :462-468
[2]   DIGITAL SUBTRACTION ANGIOGRAPHIC IMAGING OF CORONARY FLOW RESERVE [J].
CUSMA, JT ;
TOGGART, EJ ;
FOLTS, JD ;
PEPPLER, WW ;
HANGIANDREOU, NJ ;
LEE, CS ;
MISTRETTA, CA .
CIRCULATION, 1987, 75 (02) :461-472
[3]   DIGITAL ANGIOGRAPHIC IMPULSE-RESPONSE ANALYSIS OF REGIONAL MYOCARDIAL PERFUSION - LINEARITY, REPRODUCIBILITY, ACCURACY, AND COMPARISON WITH CONVENTIONAL INDICATOR DILUTION CURVE PARAMETERS IN PHANTOM AND CANINE MODELS [J].
EIGLER, NL ;
PFAFF, JM ;
ZEIHER, A ;
WHITING, JS ;
FORRESTER, JS .
CIRCULATION RESEARCH, 1989, 64 (05) :853-866
[4]  
GOULD K L, 1990, Journal of the American College of Cardiology, V15, p232A
[5]   EFFECTS OF CORONARY STENOSES ON CORONARY FLOW RESERVE AND RESISTANCE [J].
GOULD, KL ;
LIPSCOMB, K .
AMERICAN JOURNAL OF CARDIOLOGY, 1974, 34 (01) :48-55
[6]  
Hangiandreou N. J., 1990, THESIS U WISCONSIN M
[7]   CHARACTERIZATION OF CHANGES IN CORONARY BLOOD-FLOW DURING THE 1ST 6 SECONDS AFTER INTRACORONARY CONTRAST INJECTION [J].
HODGSON, JM ;
MANCINI, GBJ ;
LEGRAND, V ;
VOGEL, RA .
INVESTIGATIVE RADIOLOGY, 1985, 20 (03) :246-252
[8]   VALIDATION IN DOGS OF A RAPID DIGITAL ANGIOGRAPHIC TECHNIQUE TO MEASURE RELATIVE CORONARY BLOOD-FLOW DURING ROUTINE CARDIAC-CATHETERIZATION [J].
HODGSON, JM ;
LEGRAND, V ;
BATES, ER ;
MANCINI, GBJ ;
AUERON, FM ;
ONEILL, WW ;
SIMON, SB ;
BEAUMAN, GJ ;
LEFREE, MT ;
VOGEL, RA .
AMERICAN JOURNAL OF CARDIOLOGY, 1985, 55 (01) :188-193
[9]  
KIRKEEIDE RL, 1982, COMPUTER CARDIOLOGY
[10]   MEASUREMENTS OF CORONARY FLOW RESERVE - DEFINING PATHOPHYSIOLOGY VERSUS MAKING DECISIONS ABOUT PATIENT-CARE [J].
KLOCKE, FJ .
CIRCULATION, 1987, 76 (06) :1183-1189