COMPARISON OF A MODIFIED FICK METHOD WITH THERMODILUTION FOR DETERMINING CARDIAC-OUTPUT IN CRITICALLY ILL PATIENTS ON MECHANICAL VENTILATION

被引:15
作者
LYNCH, J
KAEMMERER, H
机构
[1] University Department of Anesthesiology, Köln 41, D-5000
关键词
Assumed RQ-value; Carbon dioxide elimination; Cardiac output; Modified Fick method; Thermodilution;
D O I
10.1007/BF01705160
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We compared a modified Fick method for measuring cardiac output against the thermodilution method in 11 critically-ill patients on mechanical ventilation. Oxygen consumption {Mathematical expression} was calculated indirectly, by measuring the carbon dioxide elimination {Mathematical expression} during steady state and by assuming an average respiratory quotient of 0.9. For a total of 129 measurements, the mean difference in cardiac output between both methods was 0.03±1.3 l/min (95% confidence interval, -0.19 to 0.25 l/min, p=0.7) standard deviation, with the largest differences being measured in the low cardiac output range (<5 l/min). No statistically significant difference was found between the cardiac output values obtained with either method. These data showed a good correlation between the two methods and suggest that the modified Fick method may be useful in determining cardiac output in seriously ill patients on mechanical ventilation not requiring pulmonary arterial catheterisation, or where facilities for undertaking metabolic measurements are not available. © 1990 Springer-Verlag.
引用
收藏
页码:248 / 251
页数:4
相关论文
共 26 条
[1]  
BLANCH L, 1988, INTENS CARE MED, V14, P131
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
BRAITHWAITE MA, 1968, J APPL PHYSIOL, V24, P434
[4]   CLINICAL IMPLICATIONS OF CONTINUOUS MEASUREMENT OF ENERGY-EXPENDITURE IN MECHANICALLY VENTILATED PATIENTS [J].
CARLSSON, M ;
NORDENSTROM, J ;
HEDENSTIERNA, G .
CLINICAL NUTRITION, 1984, 3 (02) :103-110
[5]   CARDIAC-OUTPUT DETERMINATION - THERMODILUTION VERSUS A NEW COMPUTERIZED FICK METHOD [J].
CARPENTER, JP ;
NAIR, S ;
STAW, I .
CRITICAL CARE MEDICINE, 1985, 13 (07) :576-579
[6]   CONTINUOUS FICK CARDIAC-OUTPUT COMPARED TO CONTINUOUS PULMONARY-ARTERY ELECTROMAGNETIC FLOW MEASUREMENT IN PIGS [J].
DAVIES, G ;
HESS, D ;
JEBSON, P .
ANESTHESIOLOGY, 1987, 66 (06) :805-809
[7]   CONTINUOUS FICK CARDIAC-OUTPUT COMPARED TO THERMODILUTION CARDIAC-OUTPUT [J].
DAVIES, GG ;
JEBSON, PJR ;
GLASGOW, BM ;
HESS, DR .
CRITICAL CARE MEDICINE, 1986, 14 (10) :881-885
[8]   CONTINUOUS, ONLINE, REAL-TIME MEASUREMENT OF CARDIAC-OUTPUT AND DERIVED CARDIORESPIRATORY VARIABLES IN THE CRITICALLY ILL [J].
HANKELN, K ;
MICHELSEN, H ;
KUBIAK, V ;
BEEZ, M ;
BOEHMERT, F .
CRITICAL CARE MEDICINE, 1985, 13 (12) :1071-1073
[9]  
HEDLEYWHYTE J, 1976, APPLIED PHYSL RESPIR, P113
[10]  
Hodges M, 1975, Crit Care Med, V3, P182, DOI 10.1097/00003246-197509000-00002