A SIMPLIFIED METHOD TO MEASURE CORONARY BLOOD-FLOW VELOCITY IN PATIENTS - VALIDATION AND APPLICATION OF A JUDKINS-STYLE DOPPLER-TIPPED ANGIOGRAPHIC CATHETER

被引:32
作者
KERN, MJ [1 ]
机构
[1] WASHINGTON UNIV,DIV CARDIOL,ST LOUIS,MO 63130
关键词
D O I
10.1016/0002-8703(90)90137-M
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To facilitate more rapid and safe measurement of coronary flow velocity reserve in patients, we developed a Judkins-style angiographic catheter tipped with a 20 MHz Doppler crystal. In 19 patients without coronary artery disease, resting and hyperemic (10 mg intracoronary papaverine) mean and phasic coronary flow velocity signals were measured with the Judkins-style and 2.5F intracoronary Doppler catheters at identical coronary loci. Mean coronary flow velocity at rest was similar (14 ± 8, 10 ± 7 cm/sec, p = ns), but was higher during hyperemia for the Judkins-Doppler (41 ± 8 versus 32 ± 14 cm/sec, p < 0.05). Coronary flow velocity reserve, calculated as the ratio of mean velocity reserve, calculated as the ratio of mean velocity at rest to mean velocity following papaverine, was 3.3 ± 1.4 and 3.7 ± 1.2 units (p = ns) for the Judkins and intracoronary Doppler techniques, respectively (r = 0.801, p < 0.001). The Judkins-style Doppler catheter technique permits flow velocity and coronary flow velocity reserve measurements that correlate strongly with those of the intracoronary catheter technique, facilitating safe, quick, and accurate assessment of coronary physiology. © 1990.
引用
收藏
页码:1202 / 1212
页数:11
相关论文
共 20 条
[1]   NEW METHOD TO MEASURE PHASIC CORONARY BLOOD VELOCITY IN MAN [J].
BENCHIMOL, A ;
STEGALL, HF ;
GARTLAN, JL .
AMERICAN HEART JOURNAL, 1971, 81 (01) :93-+
[2]   ABNORMAL CORONARY VASOMOTION DURING EXERCISE IN PATIENTS WITH NORMAL CORONARY-ARTERIES AND REDUCED CORONARY FLOW RESERVE [J].
BORTONE, AS ;
HESS, OM ;
EBERLI, FR ;
NONOGI, H ;
MAROLF, AP ;
GRIMM, J ;
KRAYENBUEHL, HP .
CIRCULATION, 1989, 79 (03) :516-527
[3]   MICROVASCULAR ANGINA AS A CAUSE OF CHEST PAIN WITH ANGIOGRAPHICALLY NORMAL CORONARY-ARTERIES [J].
CANNON, RO ;
EPSTEIN, SE .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (15) :1338-1343
[4]   PULSED DOPPLER CORONARY-ARTERY CATHETER - PRELIMINARY-REPORT OF A NEW TECHNIQUE FOR MEASURING RAPID CHANGES IN CORONARY-ARTERY FLOW VELOCITY IN MAN [J].
COLE, JS ;
HARTLEY, CJ .
CIRCULATION, 1977, 56 (01) :18-25
[5]   MULTIPLE RANGE AND MULTIPLE F TESTS [J].
DUNCAN, DB .
BIOMETRICS, 1955, 11 (01) :1-42
[6]   NONINVASIVE ASSESSMENT OF CORONARY STENOSES BY MYOCARDIAL PERFUSION IMAGING DURING PHARMACOLOGICAL CORONARY VASODILATION .8. CLINICAL FEASIBILITY OF POSITRON CARDIAC IMAGING WITHOUT A CYCLOTRON USING GENERATOR-PRODUCED RB-82 [J].
GOULD, KL ;
GOLDSTEIN, RA ;
MULLANI, NA ;
KIRKEEIDE, RL ;
WONG, WH ;
TEWSON, TJ ;
BERRIDGE, MS ;
BOLOMEY, LA ;
HARTZ, RK ;
SMALLING, RW ;
FUENTES, F ;
NISHIKAWA, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (04) :775-789
[7]   EFFECTS OF REGIONAL ALPHA-BLOCKADE AND BETA-BLOCKADE ON RESTING AND HYPEREMIC CORONARY BLOOD-FLOW IN CONSCIOUS, UNSTRESSED HUMANS [J].
HODGSON, JM ;
COHEN, MD ;
SZENTPETERY, S ;
THAMES, MD .
CIRCULATION, 1989, 79 (04) :797-809
[8]   ASSESSMENT OF SEVERITY OF CORONARY STENOSES USING A DOPPLER CATHETER - VALIDATION OF A METHOD BASED ON THE CONTINUITY EQUATION [J].
JOHNSON, EL ;
YOCK, PG ;
HARGRAVE, VK ;
SREBRO, JP ;
MANUBENS, SM ;
SEITZ, W ;
PORTS, TA .
CIRCULATION, 1989, 80 (03) :625-635
[9]   IMPAIRED CORONARY VASODILATOR RESERVE IN THE IMMEDIATE POSTCORONARY ANGIOPLASTY PERIOD - ANALYSIS OF CORONARY-ARTERY FLOW VELOCITY INDEXES AND REGIONAL CARDIAC VENOUS EFFLUX [J].
KERN, MJ ;
DELIGONUL, U ;
VANDORMAEL, M ;
LABOVITZ, A ;
GUDIPATI, CV ;
GABLIANI, G ;
BODET, J ;
SHAH, Y ;
KENNEDY, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (04) :860-872
[10]   PARADOXICAL VASOCONSTRICTION INDUCED BY ACETYLCHOLINE IN ATHEROSCLEROTIC CORONARY-ARTERIES [J].
LUDMER, PL ;
SELWYN, AP ;
SHOOK, TL ;
WAYNE, RR ;
MUDGE, GH ;
ALEXANDER, RW ;
GANZ, P .
NEW ENGLAND JOURNAL OF MEDICINE, 1986, 315 (17) :1046-1051