MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD

被引:32
作者
HAYES, DL
GRAHAM, KJ
IRWIN, M
VIDAILLET, H
DISLER, G
SWEESY, M
KINCAID, D
OSBORN, MJ
SUMAN, VJ
NEUBAUER, SA
SEEBANDT, M
KALLINEN, L
机构
[1] MAYO CLIN & MAYO FDN,BIOSTAT SECT,ROCHESTER,MN 55905
[2] MINNEAPOLIS HEART INST FDN,MINNEAPOLIS,MN
[3] ROBERT BATEY CARDIOL CTR,BRADENTON,FL
[4] GREY NUNS HOSP,EDMONTON,AB,CANADA
[5] MARSCHFIELD CLIN,DIV CARDIOL,MARSHFIELD,WI
[6] ROYAL ALEXANDRA HOSP,EDMONTON,AB,CANADA
[7] MIDELFORT CLIN,EAU CLAIRE,WI
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1995年 / 18卷 / 05期
关键词
POLYURETHANE INSULATION; PACEMAKER LEAD FAILURE; MEDTRONIC 4004/4004M PACING LEAD; PACEMAKER FOLLOW-UP;
D O I
10.1111/j.1540-8159.1995.tb04740.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A multicenter study was undertaken to determine the failure rate of a specific bipolar tined polyurethane ventricular pacing lead, the Medtronic 4004/4004M pacing lead. Seven centers in the United States and Canada implanted 586 Medtronic 4004/4004M pacing leads. The study was designed to determine the probability and clinical manifestations of lead failure. Only failures compatible with an insulation problem were included. The Kaplan-Meier estimate of the percentage of 4004/4004M lead failures within 4 years after implantation was 14.1% (95% confidence interval: 8.5%-19.3%). Failures were manifested as sensing abnormalities, failure to capture, early battery depletion, and significant decrease in measured impedance compared with previous impedance measurements. The observed rate of failure is unacceptable, and strong consideration should be given to replacing the 4004/4004M pacing lead in pacemaker dependent patients and closely monitoring nondependent patients.
引用
收藏
页码:999 / 1004
页数:6
相关论文
共 16 条
  • [1] BRINKER JA, 1991, PACE, V14, P639
  • [2] SIMULTANEOUS INSULATION DETERIORATION ASSOCIATED WITH SIDE-BY-SIDE SUBCLAVIAN PLACEMENT OF 2 POLYURETHANE LEADS
    FYKE, FE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1571 - 1574
  • [3] 16 FAILURES IN A SINGLE MODEL OF BIPOLAR POLYURETHANE-INSULATED VENTRICULAR PACING LEAD - A 44-MONTH EXPERIENCE
    HANSON, JS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03): : 389 - 394
  • [4] BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD - A 4-YEAR EXPERIENCE
    HAYES, DL
    HOLMES, DR
    MERIDETH, J
    OSBORN, MJ
    VLIETSTRA, RE
    NEUBAUER, SA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (02): : 192 - 196
  • [5] A MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    CROWSON, CS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07): : 1033 - 1039
  • [6] ANATOMICAL AND MORPHOLOGICAL EVALUATION OF PACEMAKER LEAD COMPRESSION
    JACOBS, DM
    FINK, AS
    MILLER, RP
    ANDERSON, WR
    MCVENES, RD
    LESSAR, JF
    COBIAN, KE
    STAFFANSON, DB
    UPTON, JE
    BUBRICK, MP
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 434 - 444
  • [7] NONPARAMETRIC-ESTIMATION FROM INCOMPLETE OBSERVATIONS
    KAPLAN, EL
    MEIER, P
    [J]. JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1958, 53 (282) : 457 - 481
  • [8] ANATOMICAL MECHANISMS EXPLAINING DAMAGE TO PACEMAKER LEADS, DEFIBRILLATOR LEADS, AND FAILURE OF CENTRAL VENOUS CATHETERS ADJACENT TO THE STERNOCLAVICULAR JOINT
    MAGNEY, JE
    FLYNN, DM
    PARSONS, JA
    STAPLIN, DH
    CHINPURCELL, MV
    MILSTEIN, S
    HUNTER, DW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 445 - 457
  • [9] A NEW APPROACH TO PERCUTANEOUS SUBCLAVIAN VENIPUNCTURE TO AVOID LEAD FRACTURE OR CENTRAL VENOUS CATHETER OCCLUSION
    MAGNEY, JE
    STAPLIN, DH
    FLYNN, DM
    HUNTER, DW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (11): : 2133 - 2143
  • [10] ASYMPTOTICALLY EFFICIENT RANK INVARIANT TEST PROCEDURES
    PETO, R
    PETO, J
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES A-GENERAL, 1972, 135 : 185 - &