CARDIAC FLOW MEASUREMENT BY ULTRAFAST CT - VALIDATION OF CONTINUOUS AND PULSATILE FLOW

被引:13
作者
LUDMAN, PF
DARBY, M
TOMLINSON, N
POOLEWILSON, PA
REES, S
机构
[1] Royal Brompton National Heart and Lung Hospital, London
[2] Department of Radiology, St. George’s Hospital, London
[3] South East Thames Regional Health Authority, Becks Hill-on-Sea
关键词
COMPUTED TOMOGRAPHY; CINE; (CINE-CT); PERFUSION; TECHNIQUES;
D O I
10.1097/00004728-199209000-00023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To gauge the accuracy of ultrafast CT in measuring cardiac output and myocardial perfusion in humans, measurements of continuous and pulsatile flow were made in a large asymmetrical phantom. The variation in the relationship between Hounsfield number and contrast concentration was assessed in a human thorax phantom. Radiopaque contrast medium was injected during perfusion of the phantom at a range of flow rates between 1.5 and 8 L/min. The phantom was scanned in two modes (50 and 100 ms) during continuous and pulsatile flow and with the phantom surrounded by air and by water. Flow in the tubes was calculated using indicator dilution theory, and flow in the tissue-equivalent chamber was calculated by applying first-pass distribution principles. The standard deviation of the difference between calculated and measured flow varied from 0.2 to 0.6 L/min, giving 95% limits of agreement from 0.4 to 1.2 L/min. The constant (K) relating Hounsfield unit number to iodine concentration varied widely both in different locations within the phantom and under different scan conditions (17.2-27.6 HU/mg I). Within a human thorax phantom, K varied from 14.15 to 23.18 HU/mg I and was dependent on location within the thorax phantom, the scan mode, and the cross-sectional diameter of the phantom. These data suggest that though the ultrafast CT scanner can measure continuous and pulsatile flow accurately in tubes, precise measurements of cardiac output in humans will require K to be assessed for each subject. Measurements of flow in tissue should be possible.
引用
收藏
页码:795 / 803
页数:9
相关论文
共 14 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   PRECISE MEASUREMENT OF VERTEBRAL MINERAL-CONTENT USING COMPUTED-TOMOGRAPHY [J].
CANN, CE ;
GENANT, HK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :493-500
[3]   MEASUREMENT OF CARDIAC-OUTPUT BY CINE COMPUTED-TOMOGRAPHY [J].
GARRETT, JS ;
LANZER, P ;
JASCHKE, W ;
BOTVINICK, E ;
SIEVERS, R ;
HIGGINS, CB ;
LIPTON, MJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1985, 56 (10) :657-661
[4]   MEASUREMENT OF REGIONAL MYOCARDIAL BLOOD-FLOW IN DOGS BY ULTRAFAST CT [J].
GOULD, RG ;
LIPTON, MJ ;
MCNAMARA, MT ;
SIEVERS, RE ;
KOSHOLD, S ;
HIGGINS, CB .
INVESTIGATIVE RADIOLOGY, 1988, 23 (05) :348-353
[5]   Studies on the circulation IV. Further analysis of the injection method, and of changes in hemodynamtcs under physiological and pathological conditions [J].
Hamilton, WF ;
Moore, JW ;
Kinsman, JM ;
Spurling, RG .
AMERICAN JOURNAL OF PHYSIOLOGY, 1932, 99 (03) :534-551
[6]   BLOOD-FLOW MEASUREMENTS WITH RADIONUCLIDE-LABELED PARTICLES [J].
HEYMANN, MA ;
PAYNE, BD ;
HOFFMAN, JIE ;
RUDOLPH, AM .
PROGRESS IN CARDIOVASCULAR DISEASES, 1977, 20 (01) :55-79
[7]   FLOW MEASUREMENTS WITH A HIGH-SPEED COMPUTED-TOMOGRAPHY SCANNER [J].
JASCHKE, W ;
GOULD, RG ;
ASSIMAKOPOULOS, PA ;
LIPTON, MJ .
MEDICAL PHYSICS, 1987, 14 (02) :238-243
[8]  
REITER SJ, 1987, J AM COLL CARDIOL, V9, pA161
[9]   USE OF ULTRAFAST COMPUTED-TOMOGRAPHY TO QUANTITATE REGIONAL MYOCARDIAL PERFUSION - A PRELIMINARY-REPORT [J].
RUMBERGER, JA ;
FEIRING, AJ ;
LIPTON, MJ ;
HIGGINS, CB ;
ELL, SR ;
MARCUS, ML .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (01) :59-69
[10]   REGIONAL BLOOD FLOW BY FRACTIONAL DISTRIBUTION OF INDICATORS [J].
SAPIRSTEIN, LA .
AMERICAN JOURNAL OF PHYSIOLOGY, 1958, 193 (01) :161-168