PRECISION, ACCURACY, AND IMAGE PLANE UNIFORMITY IN NMR RELAXATION-TIME IMAGING ON A 1.5-T WHOLE-BODY MR-IMAGING SYSTEM

被引:28
作者
ANDERSEN, C
JENSEN, FT
机构
[1] MR-Centre, Skejby University Hospital
[2] Department of Neurosurgery, Århus University Hospital
关键词
NMR; MAGNETIC RESONANCE IMAGING; RELAXATION TIMES; PRECISION; ACCURACY;
D O I
10.1016/0730-725X(94)92202-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In order to evaluate the accuracy, precision and image plane uniformity a phantom study was undertaken on a 1.5 T whole-body MR imaging system (Philips Gyroscan S15 HP). The manufacturers algorithm for obtaining relaxation time images was used. This method uses Ratios and Least Squares (RLSQ-algorithm) on data obtained from a combination of a multi-echo (CPMG) and an inversion recovery pulse sequence, yielding a calculated T-1-, T-2- and a proton density image simultaneously. T-1 and T-2 accuracy was measured in phantoms consisting of CuSO4 and MnCl2 in aqueous solution in different concentrations. The results were compared to a reference T-1 and T-2 obtained by spectroscopy. The accuracy error over the entire relaxation time range was expressed as the root mean square of the modulus of the difference between the reference and the image measurement and was 2.3% for T-1 and 4.1% for T-2. Precision or reproducibility was measured by II double estimations on 11 samples. The precision error was 0.35% for both T-1 and T-2. Image plane uniformity, i.e., the homogeneity of pixel values throughout the image plane, was measured by 3 consecutive image readings from a phantom of 19 cm in diam. in 8 directions 45 degrees apart and 1, 3, 5, 7, 9 cm from the center. Errors in the imaging plane were expressed as the deviation from the center value. For both T-1 and T-2 the maximum deviation was less than 5% over the entire image in the transverse plane. It is concluded that when the pulse sequence timings are carefully optimized, the mixed imaging sequence in combination with the RLSQ algorithm used in this MRI system is a reliable and precise means of obtaining relaxation time data.
引用
收藏
页码:775 / 784
页数:10
相关论文
共 21 条
  • [1] Bottomley, Hardy, Argersinger, Allen-Moore, A review of 1H magnetic resonance relaxation in pathology: Are T<sub>1</sub> and T<sub>2</sub> diagnostic, Med. Phys., 14, pp. 1-37, (1987)
  • [2] Just, Thelen, Tissue characterization with T<sub>1</sub>, T<sub>2</sub> and proton density values: Results in 160 patients with brain tumors, Radiology, 169, pp. 779-785, (1988)
  • [3] MacDonald, Bell, Smith, Kean, Tocher, Douglas, Miller, Best, Correlation of human NMR T<sub>1</sub> values measured in vivo and brain water content, Br. J. Radiol., 59, pp. 355-357, (1986)
  • [4] Barnes, McDonald, A magnetic resonance imaging study of experimental cerebral edema and its response to dexamethasone, Magn. Reson. Med., 7, pp. 125-131, (1988)
  • [5] Bell, Kean, MacDonald, Barnett, Douglass, Smith, McGhee, Miller, Tocher, Best, Brain Water measured by magnetic resonance imaging, Lancet, 329, pp. 66-69, (1987)
  • [6] Brant-Zawadzki, Bartkowski, Ortendahl, Pitts, Hylton, Nishimura, Crooks, NMR in experimental cerebral edema: Value of T<sub>1</sub> and T<sub>2</sub> calculations, A.J.N.R., 5, pp. 125-129, (1984)
  • [7] Fatouros, Inao, Marmarou, Kraft, In vivo characterization of infusion edema by magnetic resonance imaging, Advances in Neurology, pp. 533-537, (1990)
  • [8] In den Kleef, Cuppen, RLSQ: T<sub>1</sub>, T<sub>2</sub> and rho calculations, combining ratios and least squares, Magn. Reson. Med., 5, pp. 513-524, (1987)
  • [9] Tofts, du Boulay, Towards quantitative measurements of relaxation times and other parameters in the brain, Neuroradiology, 32, pp. 407-415, (1990)
  • [10] Breger, Rimm, Fischer, Papke, Haughton, T<sub>1</sub> and T<sub>2</sub> measurements on a 1.5-T commercial MR imager, Radiology, 171, pp. 273-276, (1989)