INTERACTIVE DESIGN OF MOTION-COMPENSATED GRADIENT WAVE-FORMS WITH A PERSONAL-COMPUTER SPREADSHEET PROGRAM

被引:13
作者
WENDT, RE
机构
[1] Department of Radiology, Baylor College of Medicine, Houston, Texas, 77030, One Baylor Plaza
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1991年 / 1卷 / 01期
关键词
TECHNOLOGY; GRADIENT WAVE-FORMS; MOTION COMPENSATION; PULSE SEQUENCES;
D O I
10.1002/jmri.1880010112
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A personal computer spread-sheet program was used to compute the amplitudes of the gradient pulses in motion-compensated gradient waveforms. The resulting designs for velocity-compensated, gradient-echo, frequency-encoding gradients and velocity-compensated section-select waveforms required little or no modification when implemented on two clinical magnetic resonance imagers.
引用
收藏
页码:87 / 92
页数:6
相关论文
共 4 条
  • [1] Haacke EM, Lenz GW, Improving MR image quality in the presence of motion using rephasing gradients, AJR, 148, pp. 1251-1258, (1987)
  • [2] Pattany PM, Phillips JJ, Chiu LC, Et al., Motion artifact suppression technique (MAST) for MR imaging, J Comput Assist Tomogr, 11, pp. 369-377, (1987)
  • [3] Constantinesco A, Mallet JJ, Bonmartin A, Lallot C, Briguet A, Spatial or flow velocity phase encoding gradients in NMR imaging, Magn Reson Imaging, 2, pp. 335-340, (1984)
  • [4] Yuan C, Gullberg GT, Parker DL, The solution of Bloch equations for flowing spins during a selective pulse using a finite difference method, Med Phys, 14, pp. 914-921, (1987)