TESTS OF 6 CONTINUOUS-FLOW CPAP DEVICES

被引:3
作者
SHEHABI, Y
HILLMAN, KM
NAIRN, M
BALNAVE, P
TORDA, TA
机构
[1] UNIV JORDAN,HOSP,INTENS CARE UNIT,AMMAN,JORDAN
[2] UNIV NEW S WALES,LIVERPOOL HOSP,KENSINGTON,NSW 2033,AUSTRALIA
[3] PRINCE WALES HOSP,SYDNEY,NSW,AUSTRALIA
[4] PRINCE HENRY HOSP,DEPT ANAESTHESIA,SYDNEY,NSW,AUSTRALIA
关键词
EQUIPMENT; CPAP; CONTINUOUS FLOW;
D O I
10.1177/0310057X9101900216
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Six continuous flow CPAP devices were tested for pressure fluctuation and stability of inhaled oxygen concentration under conditions of simulated respiration. Four of the systems, the Ambu, Auspap, Downs'-Vital Signs and the Drager are commercially available and two, the Prince Henry (PHH) and the Prince of Wales (POW) systems were assembled from Bird respirator parts, which were available in the respective Intensive Care Units. All appeared to be clinically effective. The Ambu offers convenience and economy of gases, the Drager showed the least pressure fluctuations. Three of the four commercial systems lacked certain safety features which could more easily be added to the two non-commercial devices.
引用
收藏
页码:237 / 243
页数:7
相关论文
共 10 条
[1]  
ANDERSEN JB, 1979, SCAND J RESPIR DIS, V60, P261
[2]  
Barach A, 1937, AM SOC CLIN INVEST, V16
[3]   PEEP AND CPAP [J].
DUNCAN, AW ;
OH, TE ;
HILLMAN, DR .
ANAESTHESIA AND INTENSIVE CARE, 1986, 14 (03) :236-250
[4]   THE DOWNS ADJUSTABLE FLOW GENERATOR [J].
FISHER, GC .
ANAESTHESIA, 1988, 43 (09) :766-769
[5]   SAFETY OF A NEW SYSTEM FOR CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
JENKINS, JG ;
VANBESOUW, JP .
ANAESTHESIA, 1986, 41 (11) :1155-1156
[6]  
PETERS RM, 1974, SURG CLIN N AM, V54, P955
[7]  
POULTON EP, 1936, LANCET, P981
[8]   CONTINUOUS POSITIVE AIRWAY PRESSURE BY FACE MASK IN ACUTE CARDIOGENIC PULMONARY-EDEMA [J].
RASANEN, J ;
HEIKKILA, J ;
DOWNS, J ;
NIKKI, P ;
VAISANEN, I ;
VIITANEN, A .
AMERICAN JOURNAL OF CARDIOLOGY, 1985, 55 (04) :296-300
[9]   LUNG-VOLUMES, MECHANICS, AND OXYGENATION DURING SPONTANEOUS POSITIVE-PRESSURE VENTILATION - THE ADVANTAGE OF CPAP OVER EPAP [J].
SCHLOBOHM, RM ;
FALLTRICK, RT ;
QUAN, SF ;
KATZ, JA .
ANESTHESIOLOGY, 1981, 55 (04) :416-422
[10]   EVALUATION OF THE AMBU CPAP SYSTEM [J].
VENN, PJH .