AN APPARATUS FOR APPLYING PURE NONCONSTRAINING MOMENTS TO SPINE SEGMENTS IN-VITRO

被引:168
作者
CRAWFORD, NR
BRANTLEY, AGU
DICKMAN, CA
KOENEMAN, EJ
机构
[1] BARROW NEUROL INST,DEPT EDITORIAL,DIV NEUROL SURG,PHOENIX,AZ 85013
[2] SPINAL BIOMECH RES LAB,PHOENIX,AZ
[3] HARRINGTON ARTHRITIS RES CTR,PHOENIX,AZ
关键词
FLEXIBILITY TESTING; SPINAL SEGMENT; TEST APPARATUS;
D O I
10.1097/00007632-199510000-00005
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This article reports on the design and use of a new apparatus for creating and monitoring pure, relatively nonconstraining moments to induce flexion/extension, lateral bending, and axial rotation in cadaveric spine segments of two or more vertebrae. Objective, The apparatus was designed to take advantage of the precision and control available in a servo-hydraulic testing frame to efficiently create and monitor testing moments. Summary of Background Data, Other laboratories have reported methods of flexility testing that also use cables and pulleys. However, instead of loading the cables and pulleys using a mechanical testing frame, previous systems have used pneumatic actuators or dead weights. Methods, Force from a uniaxial mechanical testing frame is converted to torque applied to the specimen through a system of cables and pulleys. The cable orientation is monitored to ensure that pure moments are created. Applied moments are recorded using one or two load cells, Results. Sketches of the apparatus are presented and its operation is described. Conclusion. Because the materials required to build this apparatus are inexpensive and the equipment needed for its operation is common in mechanical testing labs, this design may be useful for researchers interested in beginning in vitro spine flexibility testing with minimal expenditure.
引用
收藏
页码:2097 / 2100
页数:4
相关论文
共 5 条
[1]  
Goel V.K., Clark C.R., Gallaes K., Liu Y.K., Moment-rotation relationships of the ligamentous occipito-atlanto-axial complex, J Biomech, 21, pp. 673-680, (1988)
[2]  
Panjabi M.M., Dvorak J., Crisco J.J., Oda T., Wang P., Grob D., Effects of alar ligament transection on upper cervical spine rotation, J Orthop Res, 9, pp. 584-593, (1991)
[3]  
Panjabi M.M., Biomechanical evaluation of spinal fixation devices: 1. A conceptual framework, Spine, 13, pp. 1129-1134, (1988)
[4]  
Smith T.J., In vitro spinal biomechanics: Experimental methods and apparatus, Spine, 16, pp. 1204-1210, (1991)
[5]  
Yamamoto I., Panjabi M.M., Crisco T., Oxland T., Three- dimensional movements of the whole lumbar spine and lumbosacral joint, Spine, 14, pp. 1256-1260, (1989)