3D CORONARY MR-ANGIOGRAPHY IN MULTIPLE BREATH-HOLDS USING A RESPIRATORY FEEDBACK MONITOR

被引:85
作者
WANG, Y [1 ]
GRIMM, RC [1 ]
ROSSMAN, PJ [1 ]
DEBBINS, JP [1 ]
RIEDERER, SJ [1 ]
EHMAN, RL [1 ]
机构
[1] MAYO CLIN,ROCHESTER,MN 55905
关键词
CORONARY MAGNETIC RESONANCE ANGIOGRAPHY; CARDIAC IMAGING; RESPIRATORY MOTION; IMAGE ARTIFACTS;
D O I
10.1002/mrm.1910340104
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To reduce respiratory blur and ghosts in 3D coronary imaging, a data acquisition scheme using consistent multiple breath-holds was implemented. A navigator echo was acquired and processed in real time to dynamically measure diaphragm position. This information was provided as a visual prompt to the patient to maintain consistency in breath-hold levels such that the variation range of diastolic heart position was less than 2 mm. Preliminary results indicate that this multiple breath-hold acquisition scheme, compared with acquisition under respiration, can significantly reduce blur and ghost artifacts in 3D coronary imaging.
引用
收藏
页码:11 / 16
页数:6
相关论文
共 20 条
[1]  
EHMAN RL, 1988, RADIOLOGY, V173, P255
[2]  
FRAHM J, 1986, J COMPUT ASSIST TOMO, V10, P163
[3]   BREATH-HOLDING CAPABILITY OF ADULTS - IMPLICATIONS FOR SPIRAL COMPUTED-TOMOGRAPHY, FAST-ACQUISITION MAGNETIC-RESONANCE-IMAGING, AND ANGIOGRAPHY [J].
GAY, SB ;
SISTROM, CL ;
HOLDER, CA ;
SURATT, PM .
INVESTIGATIVE RADIOLOGY, 1994, 29 (09) :848-851
[4]   PSEUDO-GATING - ELIMINATION OF PERIODIC MOTION ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING WITHOUT GATING [J].
HAACKE, EM ;
LENZ, GW ;
NELSON, AD .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (02) :162-174
[5]   BROAD-BAND NUCLEAR MAGNETIC-RESONANCE PULSES WITH TWO-DIMENSIONAL SPATIAL SELECTIVITY [J].
HARDY, CJ ;
CLINE, HE .
JOURNAL OF APPLIED PHYSICS, 1989, 66 (04) :1513-1516
[6]   VOLUME IMAGING WITH MR PHASED-ARRAYS [J].
HAYES, CE ;
HATTES, N ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) :309-319
[7]   SPIRAL(HELICAL) CT [J].
HEIKEN, JP ;
BRINK, JA ;
VANNIER, MW .
RADIOLOGY, 1993, 189 (03) :647-656
[8]   ADAPTIVE TECHNIQUE FOR 3-DIMENSIONAL MR IMAGING OF MOVING STRUCTURES [J].
KORIN, HW ;
FELMLEE, JP ;
EHMAN, RL ;
RIEDERER, SJ .
RADIOLOGY, 1990, 177 (01) :217-221
[9]   CORONARY-ARTERIES - 3-DIMENSIONAL MR-IMAGING WITH FAT SATURATION AND MAGNETIZATION TRANSFER CONTRAST [J].
LI, D ;
PASCHAL, CB ;
HAACKE, EM ;
ADLER, LP .
RADIOLOGY, 1993, 187 (02) :401-406
[10]   MAGNETIZATION-PREPARED CARDIAC IMAGING USING GRADIENT-ECHO ACQUISITION [J].
LIU, Y ;
RIEDERER, SJ ;
EHMAN, RL .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) :271-275