ULTRASOUND CONTRAST AGENTS NUCLEATE INERTIAL CAVITATION IN-VITRO

被引:143
作者
MILLER, DL
THOMAS, RM
机构
[1] Battelle Pacific Northwest Laboratories, Richland, WA
关键词
DIAGNOSTIC ULTRASOUND; CONTRAST AGENTS; INERTIAL CAVITATION; SONOCHEMICALS; CAVITATION NUCLEATION; HYDROGEN PEROXIDE;
D O I
10.1016/0301-5629(95)93252-U
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Some ultrasound contrast agents contain stable bodies of gas, and this study was undertaken to determine if these agents could provide nuclei for inertial cavitation, Inertial cavitation was detected and assessed by the measurement of the sonochemical hydrogen peroxide after exposure to 2.17-, 2.95- or 3.8-MHz ultrasound, A noncavitating system was obtained by removing cavitation nuclei from the rotating tube exposure chambers by vacuum degassing, and from the phosphate-buffered saline medium by filtering, Albunex(R) added at 10(-2), 10(-3) or 10(-4) dilutions, or Levovist(R) added at 2 mg mL(-1), 0.2 mg mL(-1) or 0.02 mg mL(-1) all initiated significant H2O2 production for 2.17-MHz ultrasound at 0.41 MPa or higher spatial peak pressure amplitude for 5 min exposure gated at 0.25 s on and off with 60-rpm rotation, Not rotating the tube virtually eliminated H2O2 production. For 2.5-min continuous exposure, both agents initiated significant H2O2 production for 2.95-MHz exposure at 0.58 MPa or higher, but not for 3.8-MHz exposure up to 1.16 MPa. Bubble-based ultrasound contrast agents therefore appear to be able to provide nuclei for inertial cavitation in the rotating tube exposure system.
引用
收藏
页码:1059 / 1065
页数:7
相关论文
共 26 条
[1]  
APFEL RE, 1991, ULTRASOUND MED BIOL, V17, P170
[2]   THE CREVICE MODEL OF BUBBLE NUCLEATION [J].
ATCHLEY, AA ;
PROSPERETTI, A .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1989, 86 (03) :1065-1084
[3]   CHARACTERISTICS OF ALBUNEX - AIR-FILLED ALBUMIN MICROSPHERES FOR ECHOCARDIOGRAPHY CONTRAST ENHANCEMENT [J].
BARNHART, J ;
LEVENE, H ;
VILLAPANDO, E ;
MANIQUIS, J ;
FERNANDEZ, J ;
RICE, S ;
JABLONSKI, E ;
GJOEN, T ;
TOLLESHAUG, H .
INVESTIGATIVE RADIOLOGY, 1990, 25 :S162-S164
[4]   EFFECT OF A STABILIZED MICROBUBBLE ECHO CONTRAST AGENT ON HEMOLYSIS OF HUMAN ERYTHROCYTES EXPOSED TO HIGH-INTENSITY PULSED ULTRASOUND [J].
BRAYMAN, AA ;
AZADNIV, M ;
MAKIN, IRS ;
MILLER, MW ;
CARSTENSEN, EL ;
CHILD, SZ ;
RAEMAN, CH ;
MELTZER, RS ;
EVERBACH, EC .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1995, 12 (01) :13-21
[5]   ABSORPTION AND SCATTER OF ENCAPSULATED GAS FILLED MICROSPHERES - THEORETICAL CONSIDERATIONS AND SOME MEASUREMENTS [J].
DEJONG, N ;
HOFF, L ;
SKOTLAND, T ;
BOM, N .
ULTRASONICS, 1992, 30 (02) :95-103
[6]   TRANSIENT PULSATIONS OF SMALL GAS-BUBBLES IN WATER [J].
FLYNN, HG ;
CHURCH, CC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1988, 84 (05) :1863-1876
[7]  
GOLBERG BB, 1994, ULTRASOUND MED BIOL, V20, P319
[8]   INVITRO DETECTION OF CAVITATION INDUCED BY A DIAGNOSTIC ULTRASOUND SYSTEM [J].
HOLLAND, CK ;
ROY, RA ;
APFEL, RE ;
CRUM, LA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (01) :95-101
[9]   THRESHOLDS FOR TRANSIENT CAVITATION PRODUCED BY PULSED ULTRASOUND IN A CONTROLLED NUCLEI ENVIRONMENT [J].
HOLLAND, CK ;
APFEL, RE .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (05) :2059-2069
[10]  
IERNETTI G, 1971, ACUSTICA, V24, P191