PROPOSAL OF A METHOD FOR AUTOMATIC OPTIMIZATION OF LEFT-HEART ATRIOVENTRICULAR INTERVAL APPLICABLE TO DDD PACEMAKERS

被引:6
作者
CHIRIFE, R [1 ]
机构
[1] NATL ACAD MED BUENOS AIRES,POMBO FDN,INST CARDIOL,DIV ELECTROPHYSIOL & PACEMAKERS,BUENOS AIRES,DF,ARGENTINA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1995年 / 18卷 / 01期
关键词
LEFT HEART ATRIOVENTRICULAR; DDD PACEMAKERS; ATRIOVENTRICULAR OPTIMIZATION;
D O I
10.1111/j.1540-8159.1995.tb02475.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Electrical pacing of the right heart is known to cause delays in the depolarization of left heart chambers, leading to abnormal left heart AV sequence. Interatrial conduction time, defined as the time from the right atrial pacing pulse or intrinsic P to the onset of left atrial P wave, and P wave sensing delay cause a shorter left heart AV interval during atrial pacing-ventricular sensing and atrial sense-ventricular pace. Interventricular conduction time (the time from the right ventricular pacing pulse to the onset of left ventricular depolarization), lengthens left heart AV inter val during atrial sensing-ventricular pacing. These delays may add up or partly cancel out, depending on pacing mode. Thus, an algorithm for DDD pacemakers to optimize left heart AV interval by compensating for the above delays is proposed. This algorithm takes into account pacing and sensing delays to deliver a certain AV sequence to the right heart, aimed at producing a physiological left heart AV interval. The optimization of left heart AV interval is achieved by automatically changing right heart AV interval and pacing mode in accordance with known interatrial and interventricular conduction delays, and P wave sense offset.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 22 条
[1]  
Alt EU, 1987, J ELECTROPHYSIOL, V1, P250
[2]   ALTERATION OF LEFT-VENTRICULAR PERFORMANCE BY LEFT-BUNDLE BRANCH-BLOCK SIMULATED WITH ATRIOVENTRICULAR SEQUENTIAL PACING [J].
ASKENAZI, J ;
ALEXANDER, JH ;
KOENIGSBERG, DI ;
BELIC, N ;
LESCH, M .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (01) :99-104
[3]   IMPROVEMENT OF CARDIAC-FUNCTION IN PATIENTS WITH SEVERE CONGESTIVE-HEART-FAILURE AND CORONARY-ARTERY DISEASE BY DUAL-CHAMBER PACING WITH SHORTENED AV DELAY [J].
AURICCHIO, A ;
SOMMARIVA, L ;
SALO, RW ;
SCAFURI, A ;
CHIARIELLO, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (10) :2034-2043
[4]   INTERATRIAL CONDUCTION DURING CARDIAC PACING [J].
AUSUBEL, K ;
KLEMENTOWICZ, P ;
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06) :1026-1031
[5]   ALTERATIONS IN LEFT-VENTRICULAR RELAXATION DURING ATRIOVENTRICULAR PACING IN HUMANS [J].
BEDOTTO, JB ;
GRAYBURN, PA ;
BLACK, WH ;
RAYA, TE ;
MCBRIDE, W ;
HSIA, HH ;
EICHHORN, EJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (03) :658-664
[6]  
CHIRIFE R, 1975, BRIT HEART J, V37, P1281
[7]   NONPHYSIOLOGICAL LEFT HEART AV INTERVALS AS A RESULT OF DDD AND AAI PHYSIOLOGICAL PACING [J].
CHIRIFE, R ;
ORTEGA, DF ;
SALAZAR, AI .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1752-1756
[8]  
CHIRIFE R, 1990, RBM, V12, P22
[9]   PACEMAKER SYNDROME [J].
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :1-5
[10]   VENTRICULAR SEPTAL MOTION AND LEFT-VENTRICULAR DIMENSIONS DURING ABNORMAL VENTRICULAR ACTIVATION [J].
GOMES, JAC ;
DAMATO, AN ;
AKHTAR, M ;
DHATT, MS ;
CALON, AH ;
REDDY, CP ;
MORAN, HE .
AMERICAN JOURNAL OF CARDIOLOGY, 1977, 39 (05) :641-650