ARTERIAL IMAGING WITH A NEW FORWARD-VIEWING INTRAVASCULAR ULTRASOUND CATHETER .1. INITIAL STUDIES

被引:46
作者
EVANS, JL
NG, KH
VONESH, MJ
KRAMER, BL
MEYERS, SN
MILLS, TA
KANE, BJ
ALDRICH, WN
JANG, YT
YOCK, PG
ROLD, MD
ROTH, SI
MCPHERSON, DD
机构
[1] NORTHWESTERN UNIV,SCH MED,FEINBERG CARDIOVASC RES INST,CHICAGO,IL 60611
[2] NORTHWESTERN UNIV,SCH MED,DEPT PATHOL,CHICAGO,IL 60611
[3] CARDIOVASC IMAGING SYST,SUNNYVALE,CA
关键词
ULTRASONICS; IMAGING; CATHETERS;
D O I
10.1161/01.CIR.89.2.712
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Intravascular ultrasound (IVUS) of arteries is limited by the inability of current instruments to visualize beyond the catheter tip. We have developed a prototype 4-mm-diameter forward-viewing IVUS catheter (Cardiovascular Imaging Systems, Sunnyvale, Calif) that has the ability to provide B-mode cross-sectional ultrasound data for a distance of up to 2 cm distal to the catheter tip. Methods and Results To study the utility of this device, a 20-MHz forward-viewing IVUS catheter was used to examine 13 arterial segments (5 human femoral arteries, 1 human carotid artery, 7 canine arteries) in vitro and 1 phantom. After imaging, all data were compared with histology (Histo). In all cases, the IVUS catheter provided forward-viewing images corresponding to the arterial geometry and demonstrated vascular landmarks and atherosclerotic lesions. There was a good correlation between Histo-determined luminal diameters (LD) and IVUS-determined diameters for a distance of 14 mm ahead of the catheter tip: IVUS LD=1.0 Histo LD+1.3 (r=.87). Conclusions These preliminary data suggest that a forward-viewing IVUS catheter is feasible, accurate, and useful for evaluation of arterial geometry distal to the catheter tip.
引用
收藏
页码:712 / 717
页数:6
相关论文
共 16 条
[1]  
Aretz H T, 1989, Int J Card Imaging, V4, P153, DOI 10.1007/BF01745145
[2]   CORONARY-ARTERY NARROWING IN CORONARY HEART-DISEASE - COMPARISON OF CINEANGIOGRAPHIC AND NECROPSY FINDINGS [J].
ARNETT, EN ;
ISNER, JM ;
REDWOOD, DR ;
KENT, KM ;
BAKER, WP ;
ACKERSTEIN, H ;
ROBERTS, WC .
ANNALS OF INTERNAL MEDICINE, 1979, 91 (03) :350-356
[3]  
Borst C, 1989, Int J Card Imaging, V4, P127, DOI 10.1007/BF01745142
[4]   INTRAVASCULAR ULTRASOUND IMAGING - A CURRENT PERSPECTIVE [J].
COY, KM ;
MAURER, G ;
SIEGEL, RJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (07) :1811-1823
[5]   INTRACORONARY ULTRASOUND EVALUATION OF INTERVENTIONAL TECHNOLOGIES [J].
DAVIDSON, CJ ;
SHEIKH, KH ;
KISSLO, KB ;
PHILLIPS, HR ;
PETER, RH ;
BEHAR, VS ;
KONG, Y ;
KRUCOFF, M ;
OHMAN, EM ;
TCHENG, JE ;
STACK, RS .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (13) :1305-1309
[6]   INTRAVASCULAR ULTRASOUND IMAGING OF CORONARY-ARTERIES - IS 3 LAYERS THE NORM [J].
FITZGERALD, PJ ;
STGOAR, FG ;
CONNOLLY, AJ ;
PINTO, FJ ;
BILLINGHAM, ME ;
POPP, RL ;
YOCK, PG .
CIRCULATION, 1992, 86 (01) :154-158
[7]  
FREUDENBERG H, 1981, Z KARDIOL, V70, P863
[8]   VARIABLE MORPHOLOGY OF CORONARY ATHEROSCLEROSIS - CHARACTERIZATION OF ATHEROSCLEROTIC PLAQUE AND RESIDUAL ARTERIAL LUMEN SIZE AND SHAPE BY EPICARDIAL ECHOCARDIOGRAPHY [J].
MCPHERSON, DD ;
JOHNSON, MR ;
ALVAREZ, NM ;
REWCASTLE, NB ;
COLLINS, SM ;
ARMSTRONG, ML ;
KIESO, RA ;
THORPE, LJ ;
MARCUS, ML ;
KERBER, RE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (03) :593-599
[9]   DELINEATION OF THE EXTENT OF CORONARY ATHEROSCLEROSIS BY HIGH-FREQUENCY EPICARDIAL ECHOCARDIOGRAPHY [J].
MCPHERSON, DD ;
HIRATZKA, LF ;
LAMBERTH, WC ;
BRANDT, B ;
HUNT, M ;
KIESO, RA ;
MARCUS, ML ;
KERBER, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 1987, 316 (06) :304-309
[10]   ARTERIAL IMAGING WITH A NEW FORWARD-VIEWING INTRAVASCULAR ULTRASOUND CATHETER .1. 3-DIMENSIONAL RECONSTRUCTION AND DISPLAY OF DATA [J].
NG, KH ;
EVANS, JL ;
VONESH, MJ ;
MEYERS, SN ;
MILLS, TA ;
KANE, BJ ;
ALDRICH, WN ;
JANG, YT ;
YOCK, PG ;
ROLD, MD ;
ROTH, SI ;
MCPHERSON, DD .
CIRCULATION, 1994, 89 (02) :718-723