SENSORY STIMULATION BY TIME-VARYING MAGNETIC-FIELDS

被引:118
作者
COHEN, MS [1 ]
WEISSKOFF, RM [1 ]
RZEDZIAN, RR [1 ]
KANTOR, HL [1 ]
机构
[1] MASSACHUSETTS GEN HOSP,DEPT CARDIOL,BOSTON,MA 02114
关键词
D O I
10.1002/mrm.1910140226
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
When two human volunteers were imaged with magnetic field gradient dB/dt of 61 T/s RMS, the subjects reported, to our surprise, feeling muscular twitches synchronous with gradient pulses over repeated experiments. No adverse or sustained effects were seen. Experiments in a canine, intended to assess the safety of MR imaging with dB/dt of up to 66 T/s RMS, failed to induce detectable changes in the electrocardiogram or to show any signs of gross response to gradient pulsing. Although these data are preliminary, and largely anecdotal, they suggest a level above which such stimulation may occur. We believe that this is the first report of direct human stimulation in an MRI device and that determination of the stimulation threshold may have impact on the selection of appropriate operating points for magnetic imaging systems. © 1990 Academic Press, Inc. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:409 / 414
页数:6
相关论文