COMPARATIVE-STUDY OF GLASS AND ANTIMONY ELECTRODES FOR CONTINUOUS ESOPHAGEAL PH MONITORING

被引:27
作者
VANDENPLAS, Y [1 ]
HELVEN, R [1 ]
GOYVAERTS, H [1 ]
机构
[1] JANSSEN PHARMACEUT, B-2340 BEERSE, BELGIUM
关键词
D O I
10.1136/gut.32.6.708
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Oesophageal pH monitoring data are influenced by numerous patient or technique related factors, or both. In this study, data recorded with glass microelectrodes or antimony electrodes, both with external cutaneous Ag/AgCl reference electrodes, connected to two different recording devices are compared. Pearson correlation coefficients regarding the reflux index (% of the investigation time with a pH < 4) ranged from poor to excellent (0.55 to 1.00). Differences in data recorded with glass electrodes are not significantly dependent on the recording device (r:0.93 to 1.00 for the reflux index). The correlation between data simultaneously recorded with antimony electrodes is lower, although not significantly different for the reflux index (0.90). Correlation between data recorded with glass and antimony probes is rather poor (0.55 to 0.70). Therefore data recorded with antimony electrodes cannot be compared with those recorded with glass electrodes. The mean pH recorded with glass electrodes is lower than that with antimony probes.
引用
收藏
页码:708 / 712
页数:5
相关论文
共 23 条
[1]  
ANDERSON J, 1988, GAQSTROENTEROL IN S1, V1
[2]   ON THE USE OF MONOCRYSTALLINE ANTIMONY PH ELECTRODES IN GASTRO-ESOPHAGEAL FUNCTIONAL-DISORDERS [J].
ASK, P ;
EDWALL, G ;
JOHANSSON, KE ;
TIBBLING, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1982, 20 (03) :383-389
[3]  
BEHOLZ S, 1988, GASTROENTEROL INT S1, V1
[4]   PH MEASUREMENT IN THE ESOPHAGUS [J].
BENNETT, JR .
BAILLIERES CLINICAL GASTROENTEROLOGY, 1987, 1 (04) :747-767
[5]  
BLAND JM, 1988, LANCET, P307
[6]  
DECAESTECKER JS, 1989, NETH J MED, V34, pS20
[7]  
DESVARANNES SB, 1987, GASTROEN CLIN BIOL, V11, P382
[8]  
EMDE C, 1989, NETH J MED, V34, pS3
[9]   TECHNICAL ASPECTS OF INTRALUMINAL PH-METRY IN MAN - CURRENT STATUS AND RECOMMENDATIONS [J].
EMDE, C ;
GARNER, A ;
BLUM, AL .
GUT, 1987, 28 (09) :1177-1188
[10]  
GIGNOUX C, 1987, GASTROEN CLIN BIOL, V11, P17