ACCURACY AND PERFORMANCE OF THE TERUMO ES-H51, A NEW PORTABLE BLOOD-PRESSURE MONITOR

被引:9
作者
IMAI, Y
HASHIMOTO, J
MINAMI, N
MUNAKATA, M
WATANABE, N
SAKUMA, H
SEKINO, H
ABE, K
机构
[1] Second Department of Medicine, Tohoku University School of Medicine, Kojinkai Hospital (H. Sekino), Sendai
关键词
SPHYGMOMANOMETER; KOROTKOFF SOUND METHOD; CUFF-OSCILLOMETRIC METHOD; PORTABLE;
D O I
10.1093/ajh/7.3.255
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
A high performance portable automatic sphygmomanometer, the Terumo ES-H51 (104 g 58 x 22 x 92 mm), was newly developed for clinical use as a substitute for auscultation using a mercury sphygmomanometer. This device usually displays blood pressure (BP) values obtained by the Korot-koff sound method (K-method). However, when the device judges that BP values obtained by the K-method are inaccurate or unreliable, it substitutes automatically BP values obtained by the cuff-oscillometric method (O-method). The accuracy and reliability of the device was tested by comparing it to the auscultation with the standard mercury sphygmomanometer. The mean difference between BP values obtained by the standard method and those obtained by the K-method were -0.7 +/- 2.9 mm Hg systole (mean +/- SD) and -0.3 +/- 2.6 mm Hg diastole, whereas the difference between the former and those obtained by the O-method were 0.3 +/- 5.7 mm Hg systole and 0.3 +/- 4.3 mm Hg diastole (n = 170). The agreement between the BP values obtained according to each of the two methods using the device and the standard method was within 5 mm Hg for 72% to 93% of both systolic and diastolic readings. Therefore, BP values measured by the ES-H51 are accurate. The ES-H51 is sufficiently small and light to be carried easily anywhere. The objective and reproducible BP information obtained by the present device would be useful in clinical practice.
引用
收藏
页码:255 / 260
页数:6
相关论文
共 6 条
[1]  
Kaplan N.M., Hypertension in the individual patient, Clinical Hypertension, pp. 26-53, (1990)
[2]  
Imai Y., Abe K., Sasaki S., Et al., Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: Comparison between cuff-oscillometric and microphone methods, J Hypertens, 7, pp. 983-990, (1987)
[3]  
Imai Y., Sasaki S., Minami N., Et al., The accuracy and performance of the A & D TM 2421—a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkoff sound technique, Am J Hypertens, 5, pp. 719-726, (1992)
[4]  
O'Brien E., Petrie J., Littler W., Et al., The British Hypertension Society protocol for the evaluation of automated and semi-automatic blood pressure measuring devices with special reference to ambulatory systems, J Hypertens, 8, pp. 607-619, (1990)
[5]  
(1987)
[6]  
Bland J., Altman D.G., Statistical methods for assessing agreement between two methods of clinical measurement, Lancet, 1, pp. 307-310, (1986)