EXTRACTION OF IMPLANTED TRANSVENOUS PACING LEADS - A REVIEW OF A PERSISTENT CLINICAL PROBLEM

被引:61
作者
MYERS, MR
PARSONNET, V
BERNSTEIN, AD
机构
[1] HUNTINGTON HOSP, PACEMAKER CTR, PASADENA, CA 91105 USA
[2] UNIV MED & DENT NEW JERSEY, DEPT SURG, NEWARK, NJ 07103 USA
[3] NEWARK BETH ISRAEL MED CTR, PACEMAKER CTR, NEWARK, NJ 07112 USA
[4] NEWARK BETH ISRAEL MED CTR, DEPT SURG, NEWARK, NJ 07112 USA
关键词
D O I
10.1016/0002-8703(91)90203-T
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Within a few months of implantation, permanent pacemaker leads become ensheathed in fibrocollagenous tissue. This tissue may anchor the lead so that it is difficult, dangerous, or impossible to remove it. Leads with bulbous or finned tips are particularly resistant to extraction. The risks of applying traction to an entrapped lead include induction of bradycardia or ventricular tachycardia and fibrillation, invagination of the right ventricle, avulsion of the right ventricular myocardium or tricuspid valve, hemopericardium, and cardiac tamponade. Forceful traction may result in uncoiling of the conductor, disruption of the insulation, or complete fracture, leaving an intravascular remnant that may embolize or be a source for thrombosis. Although fixation and abandonment of an inactive chronically implanted lead is frequently appropriate and is known to pose little long-term risk, the retained inactive lead may interact adversely with a new active lead and then increase the risk of venous thrombosis, serve as a potential nidus for infection, or produce spurious electrical sensing signals that may be sensed by the pulse generator. Absolute indications for lead removal are those in which there would be a life-threatening situation if the lead were to remain in situ. In the absence of an absolute indication, the decision to proceed with extraction must be made by weighing the potential for serious morbidity or mortality against the risks of the extraction technique. Techniques for lead removal include traction and open cardiotomy operations. When a portion of the lead is intravascular, forceps, snares, baskets, countertraction, or lead-transection devices may be used to retrieve the fragment. © 1991.
引用
收藏
页码:881 / 888
页数:8
相关论文
共 93 条
[1]  
[Anonymous], COMMUNICATION
[2]  
AUGE JM, 1984, CATHETER CARDIO DIAG, V10, P625, DOI 10.1002/ccd.1810100616
[3]   RETRIEVAL OF LOST CATHETERS BY A MYOCARDIAL BIOPSY CATHETER DEVICE [J].
BASHOUR, TT ;
BANKS, T ;
CHENG, TO .
CHEST, 1974, 66 (04) :395-396
[4]   PAPILLARY MUSCLE RUPTURE OF MITRAL-VALVE COMPLICATING REMOVAL OF A PERMANENT TRANSVENOUS ELECTRODE [J].
BATTLE, WE ;
BANAS, JS ;
LEVINE, HJ .
CHEST, 1973, 63 (03) :455-457
[5]  
BIGUTAY AM, 1968, JAMA-J AM MED ASSOC, V203, P176
[6]   INCARCERATION OF TRANSVENOUS PACEMAKER ELECTRODE . REMOVAL BY TRACTION [J].
BILGUTAY, AM ;
JENSEN, NK ;
SCHMIDT, WR ;
GARAMELLA, JJ ;
LYNCH, MF .
AMERICAN HEART JOURNAL, 1969, 77 (03) :377-+
[7]   TECHNIQUES OF NONSURGICAL RETRIEVAL OF LATROGENIC FOREIGN BODIES FROM HEART [J].
BLOOMFIELD, DA .
AMERICAN JOURNAL OF CARDIOLOGY, 1971, 27 (05) :538-+
[8]  
BRODOF J, 1981, AM HEART J, V104, P1112
[9]  
BYRD CL, 1985, J THORAC CARDIOV SUR, V89, P142
[10]  
BYRD CL, 1987, PACE, V10, P654