STIMULATION CHARACTERISTICS OF A STEROID-ELUTING ELECTRODE COMPARED WITH 3 CONVENTIONAL ELECTRODES

被引:30
作者
KLEIN, HH
STEINBERGER, J
KNAKE, W
机构
[1] UNIV MARBURG,KLINIKUM LAHNBERGE,DEPT CARDIOL,W-3550 MARBURG,GERMANY
[2] UNIV GOTTINGEN,DEPT CARDIOL,W-3400 GOTTINGEN,GERMANY
[3] UNIV HOSP GOTTINGEN,DEPT CARDIOTHORAC SURG,GOTTINGEN,GERMANY
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 02期
关键词
steroid‐elufing electrodes;
D O I
10.1111/j.1540-8159.1990.tb05061.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Stimulation thresholds of a steroid‐eluting electrode (Medtronic CapSure 4003, 9 patients) were compared with three conventional electrodes (carbon‐tipped lead, Siemens‐Elema 423 S/60, 10 patients; Elgiloy‐tipped lead, Cordis Encor, 10 patients; platinum‐tipped lead, Telectronix Laserdish 030–276, 9 patients). Voltage thresholds were determined during implantation, 1–3 days, 6 weeks, and 6 months postimplantation. No significant difference among the four electrodes was found in regard to stimulation and sensing behavior during implantation. Compared to intraoperative measurements, mean increase in voltage threshold and absolute voltage thresholds were substantially Jess for the steroid‐eluting electrode than for the other tested electrodes 6 weeks and 6 months after implantation. The rheobasechronaxie product, a parameter of the stimulation performance of electrodes, underlined the superior pacing characteristics of the steroid‐eluting electrode. To achieve low chronic stimulation threshold by pharmacological means is an attractive direction for future electrode technology. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:134 / 137
页数:4
相关论文
共 15 条
  • [1] BAROLD SS, 1981, PROG CARDIOVASC DIS, V24, P1
  • [2] CELLULAR ELECTROPHYSIOLOGIC EFFECTS OF DEXAMETHASONE SODIUM-PHOSPHATE - IMPLICATIONS FOR CARDIAC STIMULATION WITH STEROID-ELUTING ELECTRODES
    BENDITT, DG
    KRIETT, JM
    RYBERG, C
    GORNICK, CC
    STOKES, KL
    DETLOFF, BLS
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1989, 22 (01) : 67 - 73
  • [3] BINNER L, 1985, PACEMAKER LEADS, P371
  • [4] BUCKING J, 1984, DEUT MED WOCHENSCHR, V17, P671
  • [5] FORBATH PG, 1987, PACE, V10, P676
  • [6] HEINEMANN F, 1988, European Heart Journal, V9, P268
  • [7] COMPARISON OF PACING ELECTRODES OF DIFFERENT SHAPE AND MATERIAL - RECOMMENDATIONS
    IRNICH, W
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02): : 422 - 426
  • [8] THE CHRONAXIE TIME AND ITS PRACTICAL IMPORTANCE
    IRNICH, W
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (03): : 292 - 301
  • [9] EXPERIENCE WITH TELEMETRY-SUPPORTED PACEMAKER CONTROL IN PATIENTS WITH VVI-PACEMAKERS
    KLEIN, HH
    HANH, BK
    HELLBERG, K
    RUSCHEWSKI, W
    DEVIVIE, ER
    KREUZER, H
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (38) : 1447 - 1451
  • [10] LONG-TERM PERFORMANCE OF ENDOCARDIAL LEADS WITH STEROID-ELUTING ELECTRODES
    KRUSE, IM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06): : 1217 - 1219