PULSE RISETIME DISCRIMINATION FOR 3HE COUNTERS

被引:11
作者
CUTTLER, JM
GREENBERGER, S
SHALEV, S
机构
[1] Department of Nuclear Science, Technion - Israel Institute of Technology, Haifa
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 75卷 / 02期
关键词
D O I
10.1016/0029-554X(69)90614-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Risetime discrimination is discussed from the point of view of optimizing the experimental conditions of a 3He gas counter for fast neutron spectroscopy. The approach is to transform the pulse risetime into a crossover time (which is easily transformed to an amplitude and stored in a multichannel analyzer), using a double RC differentiating circuit and a crossover pickoff. It was required to maximize accuracy and sensitivity. The behavior of the pertinent factors (attenuation, slope at crossover, sensitivity) was determined as a function of the differentiating time constants T1 and T2, and the pulse risetime constant τ. The relationships T1≧2τ and T2≧5τ are shown to represent an optimized system. © 1969.