EVALUATION OF THE SEPTI-CHEK AFB SYSTEM IN THE RECOVERY OF MYCOBACTERIA

被引:17
作者
WHITTIER, PS [1 ]
WESTFALL, K [1 ]
SETTERQUIST, S [1 ]
HOPFER, RL [1 ]
机构
[1] UNIV N CAROLINA,DEPT MICROBIOL & IMMUNOL,CHAPEL HILL,NC 27514
关键词
D O I
10.1007/BF01962373
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The performance of the Septi-Chek AFB System (Roche) in the isolation of mycobacteria was compared to that of culture on Lowenstein-Jensen (LJ) medium and the Bactec radiometric system. The Septi-Chek AFB system detected a significantly higher number of positive specimens (62/66 versus 47/66 for Bactec and 39/56 for LJ medium) and was more often the only medium in which an isolate was recovered. The average time for detection of isolates was very similar for the Septi-Chek AFB and Bactec systems which were both significantly faster than LJ medium in the majority of isolates.
引用
收藏
页码:915 / 918
页数:4
相关论文
共 4 条
  • [1] EVALUATION OF THE ROCHE SEPTI-CHEK AFB SYSTEM FOR RECOVERY OF MYCOBACTERIA
    DAMATO, RF
    ISENBERG, HD
    HOCHSTEIN, L
    MASTELLONE, AJ
    ALPERSTEIN, P
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1991, 29 (12) : 2906 - 2908
  • [2] Fleiss J. L., 1981, STATISTICAL METHODS, P114
  • [3] COLLABORATIVE FEASIBILITY STUDY OF A BIPHASIC SYSTEM (ROCHE SEPTI-CHEK AFB) FOR RAPID DETECTION AND ISOLATION OF MYCOBACTERIA
    ISENBERG, HD
    DAMATO, RF
    HEIFETS, L
    MURRAY, PR
    SCARDAMAGLIA, M
    JACOBS, MC
    ALPERSTEIN, P
    NILES, A
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1991, 29 (08) : 1719 - 1722
  • [4] Roberts GD, 1991, MANUAL CLIN MICROBIO, P304