COMPARISON OF PULSE OXIMETERS - ACCURACY AT LOW ARTERIAL-PRESSURE IN VOLUNTEERS

被引:36
作者
FALCONER, RJ
ROBINSON, BJ
机构
[1] WELLINGTON SCH MED,DEPT SURG,ANAESTHESIA SECT,POB 7343,WELLINGTON,NEW ZEALAND
[2] WELLINGTON HOSP,DEPT ANAESTHESIA,WELLINGTON,NEW ZEALAND
关键词
Equipment: pulse oximeters; Monitoring: pulse oximeters; arterial pressure;
D O I
10.1093/bja/65.4.552
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A laboratory model was developed of limb hypoperfusion in volunteers, using partial occlusion of the brachial artery with consequent reduction in radial artery pulse pressure. This was used to compare the function of 13 pulse oximeters and the effect of reduced pulse pressure and mild hypoxia on these devices. With the exception of one device, all the pulse oximeters studied demonstrated similar accuracies at pulse pressures exceeding 20 mm Hg. There were however, significant differences between several of the pulse oximeters in both ability to display readings and accuracy of readings displayed when brachial artery occlusion reduced radial artery pulse pressures equal to or less than 20 mm Hg. © 1990 Copyright: 1990 British Journal of Anaesthesia.
引用
收藏
页码:552 / 557
页数:6
相关论文
共 11 条