ENHANCEMENT OF THE RESPONSE OF A LACTATE OXIDASE/PEROXIDASE-BASED FIBEROPTIC SENSOR BY COMPARTMENTALIZATION OF THE ENZYME LAYER

被引:29
作者
BERGER, A [1 ]
BLUM, LJ [1 ]
机构
[1] UNIV LYON 1,GENIE ENZYMAT LAB,CNRS,URA 1535,F-69622 VILLEURBANNE,FRANCE
关键词
BIOSENSOR; CHEMILUMINESCENCE; FIBEROPTIC SENSOR; FLOW INJECTION ANALYSIS; ENZYMATIC COMPARTMENTALIZED SYSTEM; LACTATE; LUMINOL;
D O I
10.1016/0141-0229(94)90008-6
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A bienzyme fiberoptic sensor for the chemiluminescent flow injection analysis of L-lactate was developed. Lactate oxidase and peroxidase were covalently immobilized on preactivated polyamide membrane either separately on different membranes or randomly coimmobilized on the same membrane. Hydrogen peroxide, generated by the lactate oxidase reaction in the presence of L-lactate, was the substrate of the second reaction catalyzed peroxidase, in the course of which light was emitted in the presence of luminol. Compartmentalization of the enzyme layer was obtained by stacking a peroxidase membrane on a lactate oxidase membrane at the sensing tip of the fiberoptic sensor. When using such a sensing tip, the response of the fiberoptic sensor for lactate was 22 times higher than the response obtained when using a membrane bearing the two enzymes randomly coimmobilized. With the compartmentalized sensor tip, the detection limit was 250 pmol lactate and the coefficient of variation was 1.7% for 10 replicates of 6.25 nmol lactate. The method was applied for lactate determination in reconstituted whey solutions, and the results were in good agreement with the analysis by an enzyme electrode.
引用
收藏
页码:979 / 984
页数:6
相关论文
共 17 条
[1]  
ASSOLANTVINET CH, 1986, ANAL LETT, V19, P875
[2]   A RELIABLE L-LACTATE ELECTRODE WITH A NEW MEMBRANE FOR ENZYME IMMOBILIZATION FOR AMPEROMETRIC ASSAY OF LACTATE [J].
BARDELETTI, G ;
SECHAUD, F ;
COULET, PR .
ANALYTICA CHIMICA ACTA, 1986, 187 :47-54
[3]  
Bardeletti G., 1991, BIOSENSOR PRINCIPLES, P7
[4]   CONTINUOUS-FLOW BIOLUMINESCENT ASSAY OF NADH USING A FIBEROPTIC SENSOR [J].
BLUM, LJ ;
GAUTIER, SM ;
COULET, PR .
ANALYTICA CHIMICA ACTA, 1989, 226 (02) :331-336
[5]   AMPEROMETRIC SENSOR AND IMMOBILIZED ENZYME ELECTRODE FOR THE DETERMINATION OF ENZYMATIC-ACTIVITIES [J].
BLUM, LJ ;
BERTRAND, C ;
COULET, PR .
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS, 1983, 16 (07) :525-540
[6]   CHEMILUMINESCENT FLOW-INJECTION ANALYSIS OF GLUCOSE IN DRINKS WITH A BIENZYME FIBEROPTIC BIOSENSOR [J].
BLUM, LJ .
ENZYME AND MICROBIAL TECHNOLOGY, 1993, 15 (05) :407-411
[7]  
BLUM LJ, 1991, FIBER OPTIC CHEM SEN, V2, P301
[8]  
BLUM LJ, 1991, BIOSENSOR PRINCIPLES, P213
[9]   PRACTICAL ASPECTS OF COUPLING ENZYME THEORY [J].
BROOKS, SPJ ;
SUELTER, CH .
ANALYTICAL BIOCHEMISTRY, 1989, 176 (01) :1-14
[10]  
COULET PR, 1990, PRACTICAL FLUORESCEN, P543