MATRIX INTERFERENCES OBSERVED WITH A THERMOSPRAY SAMPLE INTRODUCTION SYSTEM FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:24
作者
DELOOSVOLLEBREGT, MTC
PENG, RZ
TIGGELMAN, JJ
机构
[1] Laboratory of Analytical Chemistry, Delft University of Technology, 2628 RZ Delft
[2] Center of Geological Analysis, Jiangxi Bureau of Geology and Mineral Resources, Nanchang
关键词
INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETRY; THERMOSPRAY SAMPLE INTRODUCTION; MATRIX INTERFERENCE; V-GROOVE NEBULIZER;
D O I
10.1039/ja9910600165
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Matrix interferences due to the presence of metallic ions were studied for a thermospray sample introduction system used with inductively coupled plasma atomic emission spectrometry. Corresponding results obtained with a V-groove nebulizer indicated that the interferences were more pronounced in the thermospray system. In the presence of mineral acids (HCI, HCIO4 and H3PO4), similar interferences were observed for the thermospray sample introduction system and the V-groove nebulizer. The excitation temperature of the plasma was about 500 K lower when the sample was introduced in the form of a thermospray. The electron density was also lower when using the thermospray system.