SYNTHESIZING AND FITTING LINEAR POSITION-SENSITIVE DETECTOR STEP-SCANNED LINE-PROFILES

被引:38
作者
CHEARY, RW
COELHO, A
机构
关键词
D O I
10.1107/S0021889893014165
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
An X-ray powder diffractometer has been modified by the addition of a narrow-bandpass germanium pre-monochromator and a linear position-sensitive detector (LPSD) with its centre set at the normal theta-2 theta focusing condition. Diffraction data are recorded using a step-scan procedure in which the patterns recorded at each step are summed to form the final pattern over a wide angular range. In this way, diffraction patterns covering 120 degrees 2 theta can be recorded 100 times more rapidly than with conventional receiving-slit diffractometers for the same level of counting statistics. In this paper, an analysis has been carried out of the contribution of the instrumental parameters to the shapes of the X-ray diffraction lines obtained with either a stationary or a step-scanned LPSD. This is done by calculating theoretical profiles for defocusing, parallax error, thermal-noise broadening and LPSD pixel size and convoluting them with the Cu K alpha(1) emission spectrum and aberration profiles associated with a standard focusing powder diffractometer. Theoretical profiles for fitting to experimental step-scan data are synthesized by summing convoluted profiles across the detector window. The validity of this procedure is tested by fitting to experimental step-scan data from well characterized reference specimens of MgO and Y3Al5O12 (YAG). The extra broadening associated with the inclusion of a LPSD is refined in terms of the depth of the detector, the angular window of the LPSD and the angle of divergence of the incident beam. Good fits have been obtained to the reference profiles and the physical parameters of the diffractometer determined in this way agree well with directly measured values.
引用
收藏
页码:673 / 681
页数:9
相关论文
共 14 条
[1]   STUDY OF THE K-ALPHA EMISSION-SPECTRUM OF COPPER [J].
BERGER, H .
X-RAY SPECTROMETRY, 1986, 15 (04) :241-243
[2]   DESIGN AND PROPERTIES OF POSITION-SENSITIVE PROPORTIONAL-COUNTERS USING RESISTANCE-CAPACITANCE POSITION ENCODING [J].
BORKOWSKI, CJ ;
KOPP, MK .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (08) :951-962
[3]   RECENT IMPROVEMENTS TO RC-LINE ENCODED POSITION-SENSITIVE PROPORTIONAL-COUNTERS [J].
BORKOWSKI, CJ ;
KOPP, MK .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1978, 11 (OCT) :430-434
[4]   A FUNDAMENTAL PARAMETERS APPROACH TO X-RAY LINE-PROFILE FITTING [J].
CHEARY, RW ;
COELHO, A .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1992, 25 (pt 2) :109-121
[5]  
GOBEL HE, 1981, ADV XRAY ANAL, V24, P123
[6]  
GOBEL HE, 1979, ADV XRAY ANAL, V22, P255
[7]   PROFILE AGREEMENT INDEXES IN RIETVELD AND PATTERN-FITTING ANALYSIS [J].
HILL, RJ ;
FISCHER, RX .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1990, 23 :462-468
[8]   GEOMETRICAL PROBLEMS WITH A POSITION-SENSITIVE DETECTOR EMPLOYED ON A DIFFRACTOMETER, INCLUDING ITS USE IN THE MEASUREMENT OF STRESS [J].
JAMES, MR ;
COHEN, JB .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1979, 12 (AUG) :339-345
[9]  
KOPP M, 1989, MANUFACTURERS INFORM
[10]   PEAK SHAPE AND RESOLUTION IN CONVENTIONAL DIFFRACTOMETRY WITH MONOCHROMATIC X-RAYS [J].
LOUER, D ;
LANGFORD, JI .