APPLICATION OF CATHETER-TIP ISFET FOR CONTINUOUS INVIVO MEASUREMENT

被引:75
作者
SHIMADA, K
YANO, M
SHIBATANI, K
KOMOTO, Y
ESASHI, M
MATSUO, T
机构
[1] OKAYAMA UNIV,INST THERMAL SPRING RES,DEPT SURG,MISASA,TOTTORI,JAPAN
[2] TOHOKU UNIV,DEPT ELECT ENGN,SENDAI,MIYAGI 980,JAPAN
关键词
CARBON DIOXIDE - Pressure Measurement - TRANSISTORS; FIELD EFFECT - Applications;
D O I
10.1007/BF02441899
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A catheter-tip pH sensor was developed for possible clinical and biomedical application, using an ion-sensitive field-effect transistor (i. s. f. e. t. ). Its signal drift in buffer solutions during continuous measurement was less than 3 mV in 24 h (equivalent to about 0. 05 pH in 24 h). The electric insulation was perfect when soaked in buffer solutions for more than 3 months. A continuous in vivo pH measurement was successfully achieved in nonheparinised arterial blood of a Japanese monkey. There was no significant sign of clot formation for 8 hours. It revealed quick responses to respiratory acidosis and alkalosis, and the observed pH values were in accordance with those confirmed by a blood pH analyzer. Also, a catheter-tip pCO//2 sensor was prepared, using the catheter-tip pH i. s. f. e. t. The response time was 2 min for a 90% change in CO//2 tension. The muscle pCO//2 monitoring tested with it indicated a similar result to those with medical mass spectrometer.