REUSE OF BALLOON CATHETERS FOR CORONARY ANGIOPLASTY - A POTENTIAL COST-SAVING STRATEGY

被引:41
作者
PLANTE, S
STRAUSS, BH
GOULET, G
WATSON, RK
CHISHOLM, RJ
机构
[1] CHU SHERBROOKE, DIV CARDIOL, SHERBROOKE, PQ, CANADA
[2] ST MICHAELS HOSP, DIV CARDIOL, TORONTO M5B 1W8, ON, CANADA
关键词
D O I
10.1016/0735-1097(94)90142-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This study was designed to determine the effectiveness, safety and costs associated with reuse of angioplasty catheters and to compare these results with those of a contemporary center that employed a single-use strategy. Background. Coronary angioplasty is an important but expensive procedure. To overcome the financial constraints of the Canadian health care system, reuse of angioplasty catheters is routinely practiced in some institutions. Methods. In a prospective observational study, data forms were completed after each angioplasty procedure and before patient discharge over a 10-month period. Results. A total of 693 patients underwent coronary angioplasty in the two centers. Clinical and lesion characteristics were similar except for a higher incidence of unstable angina at the reuse center (p < 0.005). The angiographic success rate was identical (88%) at both centers. The reuse center utilized more balloon catheters/lesion (mean +/- SD 2.4 +/- 1.5 vs. 1.2 +/- 0.5, p < 0.00001) and had a higher incidence of initial balloon failure (10.2% vs. 3.3%, p < 0.0001). Significant prolongation of the procedure time (81 +/- 41 vs. 68 +/- 32 min, p < 0.0001) and increased volume of contrast medium (201 +/- 86 vs. 165 +/- 61 ml, p < 0.0001) were seen in the reuse center. A higher rate of adverse clinical events (7.8% vs. 3.8%, p < 0.025) was observed in the reuse tenter, especially in patients with unstable angina. Conclusions. The reuse strategy was associated with a higher rate of adverse events, prolonged procedure time and increased use of contrast medium, especially in lesions that were not crossed by the initial balloon and in patients with unstable angina. Whether these differences are related to the reuse strategy or to differences in patient groups cannot be ascertained by this obser vational study. A multicenter randomized trial is required to further assess the safety and the cost/benefit ratio of this strategy.
引用
收藏
页码:1475 / 1481
页数:7
相关论文
共 9 条
[1]   CORONARY ANGIOPLASTY FOR UNSTABLE ANGINA [J].
DEFEYTER, PJ .
AMERICAN HEART JOURNAL, 1989, 118 (04) :860-868
[2]   PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY IN 1985-1986 AND 1977-1981 - THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTE REGISTRY [J].
DETRE, K ;
HOLUBKOV, R ;
KELSEY, S ;
COWLEY, M ;
KENT, K ;
WILLIAMS, D ;
MYLER, R ;
FAXON, D ;
HOLMES, D ;
BOURASSA, M ;
BLOCK, P ;
GOSSELIN, A ;
BENTIVOGLIO, L ;
LEATHERMAN, L ;
DORROS, G ;
KING, S ;
GALICHIA, J ;
ALBASSAM, M ;
LEON, M ;
ROBERTSON, T ;
PASSAMANI, E .
NEW ENGLAND JOURNAL OF MEDICINE, 1988, 318 (05) :265-270
[3]   PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - REPORT OF COMPLICATIONS FROM THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTE PTCA REGISTRY [J].
DORROS, G ;
COWLEY, MJ ;
SIMPSON, J ;
BENTIVOGLIO, LG ;
BLOCK, PC ;
BOURASSA, M ;
DETRE, K ;
GOSSELIN, AJ ;
GRUNTZIG, AR ;
KELSEY, SF ;
KENT, KM ;
MOCK, MB ;
MULLIN, SM ;
MYLER, RK ;
PASSAMANI, ER ;
STERTZER, SH ;
WILLIAMS, DO .
CIRCULATION, 1983, 67 (04) :723-730
[4]   THE PREVALENCE AND CLINICAL-SIGNIFICANCE OF RESIDUAL MYOCARDIAL ISCHEMIA-2 WEEKS AFTER UNCOMPLICATED NON-Q-WAVE INFARCTION - A PROSPECTIVE NATURAL-HISTORY STUDY [J].
GIBSON, RS ;
BELLER, GA ;
GHEORGHIADE, M ;
NYGAARD, TW ;
WATSON, DD ;
HUEY, BL ;
SAYRE, SL ;
KAISER, DL .
CIRCULATION, 1986, 73 (06) :1186-1198
[5]   FEVER, CHILLS, AND HYPOTENSION FOLLOWING CARDIAC-CATHETERIZATION WITH SINGLE-USE AND MULTIPLE-USE DISPOSABLE CATHETERS [J].
JACOBSON, JA ;
SCHWARTZ, CE ;
MARSHALL, HW ;
CONTI, M ;
BURKE, JP .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1983, 9 (01) :39-46
[6]   RESULTS OF PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY IN WOMEN - 1985-1986 NATIONAL HEART, LUNG, AND BLOOD INSTITUTE CORONARY ANGIOPLASTY REGISTRY [J].
KELSEY, SF ;
JAMES, M ;
HOLUBKOY, AL ;
HOLUBKOV, R ;
COWLEY, MJ ;
DETRE, KM .
CIRCULATION, 1993, 87 (03) :720-727
[7]   INCIDENCE AND CLINICAL-SIGNIFICANCE OF TRANSIENT CREATINE-KINASE ELEVATIONS AND THE DIAGNOSIS OF NON-Q-WAVE MYOCARDIAL-INFARCTION ASSOCIATED WITH CORONARY ANGIOPLASTY [J].
KLEIN, LW ;
KRAMER, BL ;
HOWARD, E ;
LESCH, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (03) :621-626
[8]   REUSE OF DISPOSABLE CATHETERS AND GUIDE WIRES [J].
RAVIN, CE ;
KOEHLER, PR .
RADIOLOGY, 1977, 122 (03) :577-579
[9]  
1993, REUSE SINGLE CATHETE