METAL FILTERS FOR THE COMPENSATION OF PHOTON ENERGY-DEPENDENCE OF THE RESPONSE OF CASO4-DY - TEFLON TLD DISKS

被引:14
作者
PRADHAN, AS
BHATT, RC
机构
[1] Division of Radiological Protection, Bhabha Atomic Research Centre, Bombay, 400 085, Trombay
来源
NUCLEAR INSTRUMENTS & METHODS | 1979年 / 166卷 / 03期
关键词
D O I
10.1016/0029-554X(79)90540-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Metal filters of aluminium, stainless steel, copper, cadmium, tin and lead were studied for compensation of the photon energy dependence of the response of CaSO4:Dy Teflon TLD Discs. It was found that metals such as cadmium and tin (which are normally used for this purpose) with their K absorption edges around 30 keV, are poor compensators of photon energy dependence. However, if either cadmium or tin is used in combination with either copper or stainless steel (with the lower Z filter nearer to the TLD dosimeter disc) a better compensation of photon energy dependence is achieved. Consequently, for personnel monitoring of X and gamma radiations, a TLD badge with CaSO4:Dy Teflon TLD discs and two sets of combination filters has been recommended. In the range of 27 keV-1.25 MeV this badge offers a photon energy response which varies within ±15%. This is a considerable improvement on the corresponding limits of ±65% on using a single filter. © 1979.
引用
收藏
页码:497 / 501
页数:5
相关论文
共 16 条
[1]  
AYYANGAR K, 1974, 1ST AS REG C RAD PRO
[2]   CALCULATED ENERGY-DEPENDENCE OF SOME RTL AND RPL DETECTORS [J].
BASSI, P ;
BUSUOLI, G ;
RIMONDI, O .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1976, 27 (5-6) :291-305
[3]   ENVIRONMENTAL MONITORING WITH TLD [J].
BECKER, K .
NUCLEAR INSTRUMENTS & METHODS, 1972, 104 (02) :405-&
[4]  
DAUCH JE, 1974, 4TH P INT C LUM DOS, P651
[5]  
FEJTEK J, 1979, 4TH P INT C LUM DOS, P449
[6]   INTERNATIONAL INTERCOMPARISON OF ENVIRONMENTAL DOSIMETERS [J].
GESELL, TF ;
BURKE, GDP ;
BECKER, K .
HEALTH PHYSICS, 1976, 30 (01) :125-133
[7]  
GWIAZDOWSKI B, 1974, 4TH P INT C LUM DOS, P989
[8]  
IGA K, 1974, 4TH P INT C LUM DOS, P783
[9]  
JOHNS HE, 1971, PHYSICS RADIOLOGY, P250
[10]   EFFECTS OF PHOSPHOR PROPORTION AND GRAIN-SIZE ON PHOTON ENERGY RESPONSE OF CASO4-DY TEFLON TLD DISKS [J].
PRADHAN, AS ;
BHATT, RC .
NUCLEAR INSTRUMENTS & METHODS, 1979, 161 (02) :243-245