PRELIMINARY INVIVO BIOCOMPATIBILITY STUDIES ON PERFLUOROSULFONIC ACID POLYMER MEMBRANES FOR BIOSENSOR APPLICATIONS

被引:83
作者
TURNER, RFB
HARRISON, DJ
RAJOTTE, RV
机构
[1] UNIV ALBERTA,SURG MED RES INST,EDMONTON T6G 2N8,ALBERTA,CANADA
[2] UNIV ALBERTA,DEPT CHEM,EDMONTON T6G 2G2,ALBERTA,CANADA
[3] UNIV ALBERTA,DEPT ELECT ENGN,EDMONTON T6G 2G7,ALBERTA,CANADA
关键词
BIOSENSORS; MEMBRANES; POLYMERS; IMPLANTS;
D O I
10.1016/0142-9612(91)90003-S
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The first biocompatibility studies on the DuPont perfluorosulphonic acid polymer (Nafion) are presented. Presterilized samples of commercially cast and solution cast Nafion membranes were implanted subcutaneously, intraperitoneally and intravenously in male Sprague-Dawley rats. Scanning electron microscopy and histological examination of explanted samples and surrounding tissues reveal little, if any, evidence of acute or chronic foreign body inflammatory response. The fibrous capsules surrounding the implant remain nominally thin (< 100-mu-m) after more than 3 months in situ, while the surrounding tissue remains well vascularized. Nafion polymer exhibits sufficient biocompatibility to make it a viable candidate for some implantable biosensor applications. It may, however, be necessary to compensate for the effect of the progression of fibrous encapsulation on sensor performance, particularly during the acute response stage.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 22 条
[1]   IMPLANTED ELECTROENZYMATIC GLUCOSE SENSORS [J].
CLARK, LC ;
DUGGAN, CA .
DIABETES CARE, 1982, 5 (03) :174-180
[2]  
CZABAN JD, 1985, ANAL CHEM, V57, pA345
[3]  
EISENBERG A, 1982, ACS S SERIES, V180
[4]   NAFION-COATED ELECTRODES WITH HIGH SELECTIVITY FOR CNS ELECTROCHEMISTRY [J].
GERHARDT, GA ;
OKE, AF ;
NAGY, G ;
MOGHADDAM, B ;
ADAMS, RN .
BRAIN RESEARCH, 1984, 290 (02) :390-395
[5]   TWO-DIMENSIONAL ENZYME ELECTRODE SENSOR FOR GLUCOSE [J].
GOUGH, DA ;
LUCISANO, JY ;
TSE, PHS .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2351-2357
[6]  
GRENSHAW MA, 1989, J ELECTROCHEM SOC, V136, P414
[7]  
GUIBAULT GG, 1988, CHIMIA, V42, P267
[8]   CHARACTERIZATION OF PERFLUOROSULFONIC ACID POLYMER COATED ENZYME ELECTRODES AND A MINIATURIZED INTEGRATED POTENTIOSTAT FOR GLUCOSE ANALYSIS IN WHOLE-BLOOD [J].
HARRISON, DJ ;
TURNER, RFB ;
BALTES, HP .
ANALYTICAL CHEMISTRY, 1988, 60 (19) :2002-2007
[9]  
HUMASON GL, 1979, ANIMAL TISSUE TECHNI, P242