BEAM PROFILES ALONG THE NONWEDGED DIRECTION FOR LARGE WEDGED FIELDS

被引:11
作者
CHUI, CS
LOSASSO, T
机构
[1] Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York
关键词
TREATMENT PLANNING ALGORITHM; WEDGE PROFILES;
D O I
10.1118/1.597282
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Beam profiles along the nonwedged direction of a wedged field produced by a linear accelerator exhibit more “sagging” than that of an open field at the same depth. For large fields, the profiles of open and wedged fields can differ by as much as 1%, The extra “sagging” of wedged profiles is mainly due to the difference in penetration between on- and off-axis rays caused by the variation of beam quality across the field. An algorithm was developed to estimate an “effective” depth such that the profile of a wedged field can be approximated by the open-field profile at the effective depth. The algorithm was verified by measured beam profiles for 6- and 15-MV x-ray beams for 15°, 30°, 45°, and 60° wedges. © 1994, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1685 / 1690
页数:6