DEVELOPMENT OF A POTENTIALLY WEARABLE GLUCOSE SENSOR FOR PATIENTS WITH DIABETES-MELLITUS - DESIGN AND INVITRO EVALUATION

被引:25
作者
SCHOONEN, AJM
SCHMIDT, FJ
HASPER, H
VERBRUGGE, DA
TIESSEN, RG
LERK, CF
机构
[1] Department of Pharmaceutical Technology and Biopharmaceutics, Faculty of Pharmacy, University of Groningen, 9713 AW Groningen
关键词
enzyme perfusion; glucose oxidase; glucose sensor; oxygen electrode; subcutaneous;
D O I
10.1016/0956-5663(90)80025-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A potentially wearable glucose sensor was developed, consisting of an oxygen electrode as detector and a dynamic enzyme perfusion system as selector. The selector is a hollow fibre, which can be placed subcutaneously and dialyses glucose from tissue fluid. In this design the problems of enzyme instability and oxygen limitation might be circumvented. The sensor measures glucose reliably for over two weeks, provided a new 10 ml syringe containing a glucose oxidase solution is connected to the system each day. © 1989.