A PROSPECTIVE-STUDY UTILIZING A TRANSTELEPHONIC ELECTROCARDIOGRAPHIC TRANSMISSION PROGRAM TO MANAGE PATIENTS IN THE FIRST SEVERAL MONTHS POST-ICD IMPLANT

被引:11
作者
PORTERFIELD, JG [1 ]
PORTERFIELD, LM [1 ]
BRAY, L [1 ]
SUGALSKI, J [1 ]
机构
[1] UNIV TENNESSEE,CTR HLTH SCI,METHODIST HOSP,DEPT MED,MEMPHIS,TN 38163
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 02期
关键词
AICD; TRANSTELEPHONIC ELECTROCARDIOGRAPHIC TRANSMISSION;
D O I
10.1111/j.1540-8159.1991.tb05112.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We prospectively enrolled 20 consecutive patients (11 men and 9 women; mean age 63 +/- 9.5 years) post-AICD implant in a transtelephonic electrocardiographic transmission (TET) program. The monitor was chosen for its retrograde (30 seconds) and antegrade memory capabilities (45 seconds). The patients were discharged from the hospital after receiving instructions to utilize the system for any cardiac symptoms. The monitor was worn 1-3 months (mean 2.5 +/- 0.7 months). During the follow-up period there were 54 TETs received. Nine were for documented AICD discharges, 19 were for symptoms associated with arrhythmias (11 of these 19 reported AICD discharges that were not documented), and 26 for symptoms not associated with arrhythmias. Eight of the 9 AICD discharges documented were appropriate for ventricular tachycardia (mean 185 +/- 40 beats/min). The arrhythmias associated with symptoms were: atrial fibrillation (12); nonsustained ventricular tachycardia (3); ventricular couplets (2); ventricular premature beats (10); and atrial premature contractions (2). Several TETs documented multiple arrhythmias. The most common symptoms not associated with arrhythmias were shortness of breath, dizziness, chest pain, and nervousness. Office interrogation of the AICDs revealed 12 of the 20 patients (60%) had received AICD discharges, with 5 of these 12 patients unaware of this occurring. We found the TET monitoring system a useful tool in the management of the AICD patient the first several months postoperatively. We were able to assess device function and avoid unnecessary office visits and/or hospitalizations.
引用
收藏
页码:308 / 311
页数:4
相关论文
共 8 条
[1]  
BRODSKY AM, 1948, CIRCULATION S2, V78, P621
[2]   CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
ECHT, DS ;
ARMSTRONG, K ;
SCHMIDT, P ;
OYER, PE ;
STINSON, EB ;
WINKLE, RA .
CIRCULATION, 1985, 71 (02) :289-296
[3]   ACTUARIAL INCIDENCE AND PATTERN OF OCCURRENCE OF SHOCKS FOLLOWING IMPLANTATION OF THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
FOGOROS, RN ;
ELSON, JJ ;
BONNET, CA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (09) :1465-1473
[4]   PREDICTORS OF AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DISCHARGE FOR LIFE-THREATENING VENTRICULAR ARRHYTHMIAS [J].
KELLY, PA ;
CANNOM, DS ;
GARAN, H ;
FINKELSTEIN, D ;
MCCOMB, JM ;
MIRABAL, GS ;
ILVENTO, JP ;
RUSKIN, JN .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (01) :83-87
[5]   CLINICAL-EXPERIENCE IN 77 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
MANOLIS, AS ;
TANDEGUZMAN, W ;
LEE, MA ;
RASTEGAR, H ;
HAFFAJEE, CI ;
HUANG, SKS ;
ESTES, NAM .
AMERICAN HEART JOURNAL, 1989, 118 (03) :445-450
[6]   TIME TO 1ST SHOCK AND CLINICAL OUTCOME IN PATIENTS RECEIVING AN AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
MYERBURG, RJ ;
LUCERI, RM ;
THURER, R ;
COOPER, DK ;
ZAMAN, L ;
INTERIAN, A ;
FERNANDEZ, P ;
COX, M ;
GLICKSMAN, F ;
CASTELLANOS, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (02) :508-514
[7]  
STEINBERG JS, 1989, CIRCULATION, V80, P530
[8]   ANXIETY AND ANGER IN PATIENTS WITH VENTRICULAR TACHYARRHYTHMIAS - RESPONSES AFTER AUTOMATIC INTERNAL CARDIOVERTER DEFIBRILLATOR IMPLANTATION [J].
VLAY, SC ;
OLSON, LC ;
FRICCHIONE, GL ;
FRIEDMAN, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (02) :366-373