THE EFFECTS OF ANNEALING ON THE RESISTANCE OF HIGH-TEMPERATURE PRTS AT THE TRIPLE POINT OF WATER AFTER HIGH-TEMPERATURE USE

被引:4
作者
ANCSIN, J [1 ]
ZHANG, Y [1 ]
机构
[1] GUIZHOU PROV RES INST MEASUREMENT & TEST TECH,GUIYANG,PEOPLES R CHINA
关键词
D O I
10.1088/0026-1394/31/2/001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Experimental results indicate that R(o), the resistance of a platinum resistance thermometer at the triple point of water, is a logarithmic function of the cooling rate following high-temperature use of the thermometer. It is also shown that annealing at a constant cooling rate from the maximum temperature is a more efficient use of annealing time than soaking at constant temperatures between T-max and 450 degrees C.