TRANSCUTANEOUS LACTATE MONITORING BASED ON A MICRO-PLANAR AMPEROMETRIC BIOSENSOR

被引:21
作者
ITO, N
MATSUMOTO, T
FUJIWARA, H
MATSUMOTO, Y
KAYASHIMA, S
ARAI, T
KIKUCHI, M
KARUBE, I
机构
[1] NATL DEF MED COLL,DEPT MED ENGN,TOKOROZAWA,SAITAMA 359,JAPAN
[2] UNIV TOKYO,RCAST,MEGURO KU,TOKYO 153,JAPAN
关键词
AMPEROMETRY; LACTATE; MICRO-PLANAR; BIOSENSORS; TRANSCUTANEOUS; INTERSTITIAL FLUID;
D O I
10.1016/0003-2670(95)00216-M
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A transcutaneous lactate monitoring method has been developed by combining a suction effusion fluid (SEF) collecting technique with a micro-planar amperometric biosensor. SEF is directly collected by a weak suction through the skin from which the stratum corneum has been removed. A micro-planar biosensor, has also been developed for determining lactate concentrations in very small samples. The biosensor utilizes a lactate oxidase immobilized membrane on a micro planar platinum electrode to yield a lactate sensor. The sensor is based on amperometric detection of hydrogen peroxide generated by the lactate oxidase-catalyzed oxidation of L-lactate. The linearity of the lactate sensor response was obtained from 0.5 mM to 25 mM and the response time was about 10 s. The sensor was applied to the SEF lactate measurement of a rabbit during induced high lactate levels. Lactate concentrations measured with SEF showed good correlation with those obtained directly from whole blood. This result suggests the feasibility of transcutaneous lactate monitoring without the use of blood sampling.
引用
收藏
页码:323 / 328
页数:6
相关论文
共 23 条
[1]   AMPEROMETRIC BIOSENSOR FOR DETERMINATION OF LACTATE IN SWEAT [J].
FARIDNIA, MH ;
PALLESCHI, G ;
LUBRANO, GJ ;
GUILBAULT, GG .
ANALYTICA CHIMICA ACTA, 1993, 278 (01) :35-40
[2]  
FOXDAL P, 1992, CLIN CHEM, V38, P2110
[3]   FABRICATION AND CHARACTERIZATION OF A PLANAR ELECTROCHEMICAL-CELL AND ITS APPLICATION AS A GLUCOSE SENSOR [J].
GERNET, S ;
KOUDELKA, M ;
DEROOIJ, NF .
SENSORS AND ACTUATORS, 1989, 18 (01) :59-70
[4]   A NEEDLE-TYPE ENZYME-BASED LACTATE SENSOR FOR IN-VIVO MONITORING [J].
HU, YB ;
ZHANG, YN ;
WILSON, GS .
ANALYTICA CHIMICA ACTA, 1993, 281 (03) :503-511
[5]   DEVELOPMENT OF A TRANSCUTANEOUS BLOOD-CONSTITUENT MONITORING METHOD USING A SUCTION EFFUSION FLUID COLLECTION TECHNIQUE AND AN ION-SENSITIVE FIELD-EFFECT TRANSISTOR GLUCOSE SENSOR [J].
ITO, N ;
KAYASHIMA, S ;
KIMURA, J ;
KURIYAMA, T ;
ARAI, T ;
KIKUCHI, M ;
NAGATA, N .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (03) :242-246
[6]   SUCTION EFFUSION FLUID FROM SKIN AND CONSTITUENT ANALYSIS - NEW CANDIDATE FOR INTERSTITIAL FLUID [J].
KAYASHIMA, S ;
ARAI, T ;
KIKUCHI, M ;
NAGATA, N ;
ITO, N ;
KURIYAMA, T ;
KIMURA, J .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (05) :H1623-H1627
[7]  
KIKUCHI M, 1986, 1ST P PAN PAC S VANC, V57, P23
[8]   MINIATURE ELECTROCHEMICAL GLUCOSE BIOSENSORS [J].
KOUDELKAHEP, M ;
STRIKE, DJ ;
DEROOIJ, NF .
ANALYTICA CHIMICA ACTA, 1993, 281 (03) :461-466
[9]  
KRUSE JA, 1987, CRIT CARE CLIN, V5, P725
[10]  