POST-PLASMA-SPRAYING HEAT-TREATMENT OF THE HA COATING/TI-6AL-4V IMPLANT SYSTEM

被引:59
作者
FILIAGGI, MJ [1 ]
PILLIAR, RM [1 ]
COOMBS, NA [1 ]
机构
[1] UNIV TORONTO,CTR BIOMAT,TORONTO M5G 1G6,ONTARIO,CANADA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1993年 / 27卷 / 02期
关键词
D O I
10.1002/jbm.820270208
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The metal/ceramic interface that constitutes an important part of the plasma-sprayed HA-coated Ti-6Al-4V system may, in fact, represent the ''weak link'' in the implant design. A post-plasma-spray heat treatment to enhance chemical bonding at the metal/ceramic interface and, hence, improve the mechanical properties (interface fracture toughness and tensile coating adhesion strength) of the plasma-sprayed implant system does show promise. In preliminary heat treatment studies, however, any improvements realized were lost due to the chemical instability of the coating in a moisture-laden environment, with a concomitant loss in bonding properties. This deterioration in properties appears to be related to environmentally assisted crack growth as influenced by processing conditions. Still, an ability to improve HA/Ti-6Al-4V bonding through enhanced diffusion bonding was demonstrated, warranting further heat treatment studies involving atmosphere control during processing.
引用
收藏
页码:191 / 198
页数:8
相关论文
共 15 条
[2]   EVALUATION OF HYDROXYLAPATITE-COATED TITANIUM DENTAL IMPLANTS IN DOGS [J].
BLOCK, MS ;
KENT, JN ;
KAY, JF .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1987, 45 (07) :601-607
[3]  
Cook S D, 1987, Int J Oral Maxillofac Implants, V2, P15
[4]  
COOK SD, 1988, CLIN ORTHOP RELAT R, V232, P225
[5]   PLASMA SPRAYED COATINGS OF HYDROXYLAPATITE [J].
DEGROOT, K ;
GEESINK, R ;
KLEIN, CPAT ;
SEREKIAN, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (12) :1375-1381
[6]   STRUCTURAL-ANALYSIS OF HYDROXYAPATITE COATINGS ON TITANIUM [J].
DUCHEYNE, P ;
VANRAEMDONCK, W ;
HEUGHEBAERT, JC ;
HEUGHEBAERT, M .
BIOMATERIALS, 1986, 7 (02) :97-103
[7]   MECHANICAL AND CHEMICAL CHARACTERIZATION OF A METAL-BIOCERAMIC INTERFACE [J].
FILIAGGI, MJ ;
COOMBS, NA ;
PILLIAR, RM .
INTERFACES BETWEEN POLYMERS, METALS, AND CERAMICS, 1989, 153 :377-383
[8]   CHARACTERIZATION OF THE INTERFACE IN THE PLASMA-SPRAYED HA COATING/TI-6AL-4V IMPLANT SYSTEM [J].
FILIAGGI, MJ ;
COOMBS, NA ;
PILLIAR, RM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (10) :1211-1229
[9]  
GEESINK RGT, 1987, CLIN ORTHOP RELAT R, V225, P147
[10]   ACOUSTIC-EMISSION STUDY OF ENVIRONMENTAL-EFFECTS IN THERMALLY SPRAYED COATINGS [J].
OSTOJIC, P ;
MCPHERSON, R .
THIN SOLID FILMS, 1986, 136 (02) :215-227