MEASUREMENT OF BLOOD PERFUSION IN TISSUE USING DOPPLER ULTRASOUND

被引:85
作者
DYMLING, SO
PERSSON, HW
HERTZ, CH
机构
[1] Department of Electrical Measurements, Lund Institute of Technology, S-221 00 Lund
关键词
ULTRASONICS; DIAGNOSTIC ULTRASOUND; BLOOD PERFUSION; DOPPLER; CONTINUOUS-WAVE; ARTIFACTS;
D O I
10.1016/0301-5629(91)90179-Z
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A diagnostic tool for noninvasive evaluation of microcirculatory blood flow using continuous-wave CW Doppler ultrasound is presented. In this study, the properties of this method are investigated both theoretically and experimentally. The method utilizes a nondirectional CW Doppler flowmeter. Blood perfusion in tissue is shown to be proportional to the integral integral fS(f)df where S(f) is the Doppler power spectrum and f is the Doppler frequency. The instrumentation needed to implement the method is described. Using an experimental flow model it is demonstrated that the above integral is proportional to the product between the number of scatterers in the sample volume of the Doppler probe and the mean speed of these scatterers. This is true even for low flow velocities (down to 1 mm/s). The results from in-vivo measurments on tissues in the finger, and the calf demonstrate that the method can monitor changes in the blood perfusion. It also shows the present limitations of the method due to movement artefacts.
引用
收藏
页码:433 / 444
页数:12
相关论文
共 23 条
  • [1] ASH E, 1985, ELECTRONIC ENG, V23
  • [2] BAKER DW, 1978, ULTRASOUND ITS APPLI
  • [3] MODEL FOR LASER DOPPLER MEASUREMENTS OF BLOOD-FLOW IN TISSUE
    BONNER, R
    NOSSAL, R
    [J]. APPLIED OPTICS, 1981, 20 (12) : 2097 - 2107
  • [4] BRODY WR, 1974, IEEE T BIO-MED ENG, VBM21, P183, DOI 10.1109/TBME.1974.324381
  • [5] BRODY WR, 1971, TR49581 STANF U STAN
  • [6] CHMIEL H, 1982, Patent No. 4327739
  • [7] DYMLING SO, 1982, 5TH WORLD C ULTR MED
  • [8] DYMLING SO, 1984, 11984 LUND I TECHN D
  • [9] DYMLING SO, 1985, 31985 LUND I TECHN D
  • [10] HERTZ CH, 1981, 4TH EUR C ULTR MED D